Cleaning the Cap Piercer Video
This video provides instructions on how to properly clean the cap piercer on the Sysmex® CS-5100 System.
needed maintenance item is cleaning the piercer when residual sample is present. It's recommended the piercer be cleaned. Verify the instrument is in ready status and the light shield lid is closed. On the maintenance screen, press wipe off piercer on the operation panel. The wipe off piercer dialog box will display. Press OK. The piercer will move into position. When the piercer is in position, a second dialog box displays instructing to power off main unit. Turn off the power to the main unit. And open the light shield lid. Position the jig for wiping off under the sample probe a arm and adjust height by rotating the tabs. While holding the sample arm, wipe the piercer in an up and down motion with distilled water on gauze. Remove the jig for wiping off. Close the light shield lid. Power on the main unit. Wiping the piercer clean will be documented on the maintenance screen.
Clean instrument Rimse the Rinse Tank Replace History I Cleaning The Piercer 2012/07/19 2012/10/05 2012/07/12 2012/W22 201 OH power. Off Piercer ting after adj. a of Suøølies Item Wipe Off the piercer after turning the Win Unit Off The piercer will be mcved to wipe—off position. Wipe off Cleaning The Piercer 2012/10/0 2012/11/03 Air pressure adjustment Piercer Turn Off the pover after the execution. 2012/108/219 prime Joblist Status Mai nt. Shutdown Rermve ja m med cuvette 201 0K Ca ncel 2012/08/08 2012/09/06 Woe off piercer Lampk Lamp replacement ard calibratbn Replace fuses of the Pt»umatk: Unit Replace Piercer Syringe 2012/07/26 2012/09/11 2012/09/11 56 / 40,000 Times 21 533 / 1 21 / 1
- cs-5100 as needed maintenance
- sample arm a
- wipe off piercer