Sysmex® CS- 5100 System Multiple Dilution Analysis Job Aid
This job aid provides instructions on how to order and review Multiple Dilution Analysis results on the Sysmex CS-5100 System.
Sysmex® CS-5100 System Multiple Dilution SIEMENS Healthineers Analysis MDA Ordering 1 7 Menu Calib. Curve QC Chart Order Joblist Status Maint. Shutdown STAT Stop 1. Press Order from the IPU toolbar or menu screen. Unregistered Rack No. 000001 Change Switch 2. Select the Sample No. for which you want to set Pos. Sample No. Mc PT PTT Fbg AT DDi PTN PTTN V-INN VII-! NN Preset Detailed Settings the MDA method. 2 2 85241 Rack-Pos : -02 Sample No.: 85241 Sample Comment 4 Repeat 3 3. Press Detailed Settings on the operation panel. Reflex Assay Group Replications Dilution Ratio Measurement Time Redilution Re-analysis 4 Order Entry 4. Select the appropriate Dilution Ratio from the Micro- VII-INN 1 1/1 180 Derided 3 6 Select Dilution Ratio Settings Detailed Settings dialog box. 1/1 MDA 2 Regist 8 5. Select the appropriate MDA parameters. 9 MDA 3 MDA low3 10 Press More for additional dilution ratios. 5 MDA E MDAL More • xt. Ext. 6 [1 /17 MDA H More Ext. 1/2 (1/2) 6. Press OK from the Detailed Settings dialog box. OK Cance 1/4 1/8 PTPTT More (1/2) 7. Press Start from the IPU toolbar or main unit Cancel button. Close X CS Lab Ready 14:11 1 HOOD05162002598476 © Siemens AG 2013. All rights reserved. Page 1 Effective date: 7/17/2017 Sysmex® CS-5100 System Multiple Dilution SIEMENS Healthineers Analysis 1 Reviewing MDA Results AAM Menu Reagent Calib. Curve QC Chart Order Joblist Status Maint. Shutdown STAT Stop Start Status Sample No. Rack No. Sample Info. RVR V INN" / INN sec VII INN"% VII INN sec PTT Fbg -Pos. sec T INR sec sec mg / dl Fbs sec Order Entry 2354 000003-02 Vol N V F 17:07 1212 ***** 1. Press Joblist from the IPU toolbar or menu screen. 8824 000003-03 17-09 121.8 18.6 Browser 4 8824 000003-03 V 17:09 119.3 25.4 8824 000003-03 17.09 1325 d 335 2. View the Sample No. displaying the MDA results and Validate 85241 000003-01 08:3 116.4 18.9 the corresponding mean. 2 85241 000003-01 08:34 115.6 25.8 Condition 3 85241 000003-01 08:34 122.3 d 34.7 85241 000003-01 3. Select the line displaying the mean result. NV F 08:34 118.1 @ ***** QC CtIN503188 REAG00 M V ₣ 08:35 90.2 Find 4. Press Browser from the operation panel. Final Remeasure Hein lot A HOST No 85241 M 02/22/2017 08:26:18 5. The Browser Main screen displays the reaction curves, Validated Lip Vol # 000003-01 analysis results, and reference information from the Routine Lot Roll More selected line on the Joblist screen. 6. Press desired reaction curve graph from the Browser Main screen to display detailed information for a specific assay group on the Browser details screen. 75 VII-INN NOTE: If any errors occur during analysis, Detail flags on the VII INN sec d ***** sec 6 VII INN % 118.1 Browser Main screen. 5 25 Press Detail to find additional information about the error Testr .000 • SCr 1.000 Detail code. 5 10 50 100 2 HOOD05162002598476 © Siemens AG 2013. All rights reserved. Page 2 Effective date: 7/17/2017
- cs5100
- cs51
- 5100
- cs-5100
- sysmex
- sismex
- sysnex