Priming the Hydraulic Line Video

This video provides instructions on how to properly prime the hydraulic line on the Sysmex® CS-5100 System.

has not been in operation for an extended period of time, the hydraulic lines will need to be primed with rinse water before starting analysis. Before priming, verify there's sufficient. Rinse water to complete the process and that the waste container is empty. Open the light shield lid. Unscrew the two fixing screws of cover #1501. And remove the cover. Lift reagent arms B&C and turn them towards you. Cover the detector block with Cardboard to prevent fluid from entering the detector well. Press maintenance on the main screen. Select prime on the operation panel. A prime dialog box will display. Press OK to start the priming process. A progress bar will display during priming. When complete, the dialog box will close. Remove the Cardboard covering the detector block. Attach cover #1501 and tighten the two screws. Close the light shield lid.

a 4567 456 45. 8 9 10 11 14 21 .22.23 8 910 11 14 16 8 910 11 12 13 14 15 16 17 18 2021 22 24 8 910 11 14 15 16 8 910 11 2021 22 23 24 9 10 11 12 2021 -22 17 18 20 21 .22 2325 20 20 Priming Please wa it. Please we it. Priming hydra uk win execut«l Priming The Hydraulic Line (Execution time : approximately minutes) ure not u' ure not ure u' ure 2012/07/26 2012/07/26 55 / 55 40000 Times 55 21.466 / Times 21.466 Times Times al / 31 /