Removing Cuvette Jams Video
This video provides instructions on how to properly remove a cuvette jam on the Sysmex® CS-5100 System.
a space may form within the queue vets in the Cuvette Hopper, blocking the supply of cubits. When this occurs, the error message cuvet jammed. Small hopper will display in the error help dialog box. Press the stop alarm button. Then press the stop button on the Ipiu toolbar. An interrupt confirmation dialog box will display. Press OK. Verify instrument status is ready. Lift the light shield lid. Turn the incubator cover so the incubator catcher faces the front of the instrument. Remove the rail cover by lifting the cover diagonally and pulling forward. Insert the cuvet removal rod into the hopper and stir the Q bets. Replace the rail cover by fitting the two tabs on the left side of the rail cover into the main unit. Push the rail cover diagonally so the bottom of the rail cover fits into the groove of the main unit. Close the light shield lid. Press start on the main unit or the Ipiu toolbar when analysis is complete. Document remove jammed cuvette in manual maintenance on the operation panel.
Reagent curve ac Chart Wing Cuv Cuvette Jammed (Small Hopper) J mmed (Small Hopper) mmed (Small Ilopper) e/ Time Up Errorcode:37002 000 Stop Alarm Stop The sample rack(s) on the measurement line are ekected. 2/1 1/06 Removing A Cuvette Jam The sample rack being ek-cted. The sample rack being The sample rack is being ejected. After eh-cting the sample racKs), analysis results for button to eject the the After the Please gait until the sample rack is the jammed by using rod. Wing parameter in the not cornpketed become error. parameter in the dispensing not completed become error. error persists, Contact your error persists, Contact representative. error Contact representative. error Contact Eject the rack(s)? Log Corrective Mai nt. Error Log Error Please •a it. 13:37 000 000
- cs-5100 as needed maintenance
- cuvette jam small hopper
- cuvette removal rod