Testing Analyzer Water on the Atellica® IM Analyzer Job Aid Version 1.25 and Below

This job aid provides instructions on how to perform the as needed maintenance task of testing analyzer water for the Atellica® IM Analyzer.

Atellica® IM Analyzer SIEMENS Testing Analyzer Water Job Aid Healthineers SIEMENS .. Thineers Atellica Solution Version. 1.14,0.2983003 P-P1-04 1. Put the IM Analyzer in the Diagnostic State: A System Samples Inventory Calibration Qc Patient Orders Worklist Maintenance Setup A. On the Command bar, select System B Status Logs Operator Diagnostics About Select the Operator Diagnostics tab P1-060 B. C-P1-08 P1-060 Subsystems Diagnostic Name Description C. Select Enter Diagnostics State All Initialize Module Move all module components to their primary positions. Perform Other Clear Thermals Clear system thermal components following a thermal error that caused components to go offline. Reagent Loader Initialize Reagent Loader Return the reagent loader and the associated subsystems to the starting positions. Cuvette Loader Prime Cuvettes Cuvette conveyor moves to fill the vertical cuvette chute with cuvettes. Incubation Ring Initialize Cuvette Loader Move the cuvette loader and the associated subsystem components to their primary positions. Luminometer Empty Ring Empty all the cuvettes from the incubation ring. * This procedure is continued on the next slide. Reagent Compartment Empty and Fill Ring Simultaneously empty the cuvettes from the incubation ring and load the ring with new cuvettes. Reagent Probes Initialize Incubation Ring Move the incubation ring and the associated subsystem components to their primary positions Sample Probe Initialize Luminometer Return the luminometer and the associated subsystems to the starting positions. Tip Loader Initialize Reagent Compartment Return the reagent compartment and the associated subsystems to the starting positions. Wash Ring | Status Module Covers Parameters C Status Enter Diagnostics State Turn Mechanics Off ? PEP LabManager Wednesday, June 13, 2018 8:09 AM 1 Effective Date: 20-DEC-2020 | HOOD05162003152571 Unrestricted © Siemens Healthineers, 2020 Atellica® IM Analyzer SIEMENS Testing Analyzer Water Job Aid (continued) Healthineers 2. Open the right front door 3. Locate the water reservoir and unscrew the lid 4. Using a sterile pipette or syringe, pull out a 25-50mL sample of water 2 5. Follow laboratory protocol and test sample for microbes Note: If the bacteria count is > 50 CFU/mL, contact the local technical support provider to disinfect the water system. 6. Select System > Operator Diagnostics > Exit Diagnostic 3 State 2 Effective Date: 20-DEC-2020 | HOOD05162003152571 Unrestricted © Siemens Healthineers, 2020 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2020 Global Siemens Healthineers Headquarters Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84 0 siemens.com/healthineers 3 Effective Date: 20-DEC-2020 | HOOD05162003152571 Unrestricted © Siemens Healthineers, 2020