UniPOC™ Data Management System Add Lot Information Job Aid
This job aid provides instructions on how to manage lot information including entering reagent, control and linearity lots and their associated ranges.
UniPOC™ Data Management System SIEMENS Healthineers Add Lot Information Enter Reagent Lot Information EDIT REAGENT LOT: Main Hospital > All Departments > All Locations 1. Select Lots > Reagent 2. Locate the desired lot code and select Edit INPUT INFORMATION • If the lot code has not been recognized by the UniPOC software, users must select Add and manually enter the Reagent Type Glucose Test Strip V new lot code and reagent type Enter the Lot number * Lot Code 4ACH5H 3. 3 • The lot code and reagent type cannot be edited once a lot * Lot No. 4ACH5H code has been recognized by the UniPOC software Abbott Diabetes Care 4. Enter the additional required entry fields: In-Service Date and Manufacturer Expiration Date * In-Service Date 01/01/2021 31 5. Select the Download to Instruments checkbox to allow the Expiration Date 08/31/2021 31 4 Reagent Lot to be downloaded to instruments, if applicable • Only download reagent lots currently in use to ensure Download to Instruments instruments receive the appropriate reagent download list 5 6. Select Save 6 Save Cancel • Reagent Lot screen appears with the new lot displayed Note: It is strongly recommended to add reagent lots at the Facility level. 1 Effective Date: 12-Jul-2021 | HOOD05162003201651 Unrestricted © Siemens Healthineers, 2021 UniPOC™ Data Management System SIEMENS Healthineers Add Lot Information ADD REAGENT RANGE: Main Hospital Enter Reagent Ranges INPUT INFORMATION 1. Select a desired Lot No. checkbox 2. Select Edit Reagent Type Glucose Test Strip 3. Select Add 4. Select an analyte from the Analyte dropdown list, if Analyte GLU 4 applicable Reagent Type is a read-only field and cannot be Level • Low 5 edited 5. Select Low from the Level dropdown list 6 Range To 6. Enter the lower and upper values in the Range fields 7. Select the corresponding unit from the Unit dropdown Unit mg/dL V 7 list 8. Select Save 9. Repeat steps 3-8 to add normal and high level ranges 8 Save Cancel Note: For multi-analyte reagent entry, ranges must be entered per lot number per analyte in order to generate statistical reports. Note: Refer to vendor documentation provided with the installed instrument to determine what information is required to successfully enter reagent lot information and reagent ranges. 2 Effective Date: 12-Jul-2021 | HOOD05162003201651 Unrestricted © Siemens Healthineers, 2021 UniPOC™ Data Management System SIEMENS Healthineers Add Lot Information ADD CONTROL LOT: Main Hospital »> A Enter Control Lot Information INPUT INFORMATION 1. Select Lots > Control Select a Facility in the Tree Reagent Type Glucose Test Strip V 4 2. 3. Select Add Level Low V • Add Control Lot screen appears 4. Select a Reagent Type from the dropdown list Lot No., In-Service Date and Expiration Date are required entry * Lot No. 5. fields 6. Select Save Manufacturer • Control Lots screen appears with the new lot displayed 5 * In-Service Date 31 Note: It is strongly recommended to add control lots at the Facility level. * Expiration Date 31 Download to Instruments 6 Save Cancel 3 Effective Date: 12-Jul-2021 | HOOD05162003201651 Unrestricted © Siemens Healthineers, 2021 UniPOC™ Data Management System SIEMENS Healthineers Add Lot Information Enter Control Ranges 2 Tree Options Refresh Hide CONTROL LOTS: Main Hospital >> All Departments » All Locations 1. Select Lots > Control Organization Select a Facility, Department or Location in H Main Hospital SEARCH CRITERIA 2. DER the Tree Instrument DCA Vantage v In-Service Date 31 3 Hide Expired Lots 1 East From 3. Select an instrument and reagent type from 1 North Reagent Type DCA A/C In-Service Date V 1 South To 31 the dropdown lists 4 1 West Download All V 4. Select Display Records CLINITEK Status Plus Display Records 4 5. Select desired Lot No. checkbox A B DCA Vantage 1 Record Found A 6. Select Edit &Epoc Add Edit Delete Customize _FreeStyle Precision Pro • Edit Control Lot screen appears with A i-STAT Lot No. Manufacturer Reagent Type Level In-Service Expiration Date Download Reagent Type and Lot No. displayed Date Precision Xceed Pro IT Department 5 A 9012 DCA A/C Low -- No A @ Lab *Procedure is continued on the next slide. A Nursery Add Edit Delete Customize 6 4 Effective Date: 12-Jul-2021 | HOOD05162003201651 Unrestricted © Siemens Healthineers, 2021 UniPOC™ Data Management System SIEMENS Healthineers Add Lot Information ADD CONTROL RANGE: Main Hospital >> ER >> Enter Control Ranges (continued) INPUT INFORMATION 7. Select Add • Add Control Range screen appears Reagent Type DCA A/C V 8. Select Analyte from the dropdown list 9. Select Level from the dropdown list Analyte ALB 8 10. Enter the ranges for the selected Level • Range fields completed are dependent upon Level Level 1 9 device type. Refer to specific device’s user manual for additional information To 11. Select Unit from dropdown list Critical Range 12. Select Save Warning Range To 10 13. Repeat steps 7-12 to add additional levels and Target Value their associated ranges Unit mg/dl 11 12 Save Cancel 5 Effective Date: 12-Jul-2021 | HOOD05162003201651 Unrestricted © Siemens Healthineers, 2021 UniPOC™ Data Management System SIEMENS Healthineers Add Lot Information Enter Linearity Lot Information ADD LINEARITY LOT: Main Hospital 1. Select Lots > Linearity 2. Select a Facility in the Tree INPUT INFORMATION 3. Select an instrument and reagent type from the dropdown Reagent Type CHEM8+ lists 4. Select Display Records * Lot No. 5. Select Add • Add Linearity Lot screen appears Manufacturer 6 6. Enter linearity lot information 31 • Lot No. and Expiration Date are required entry fields In-Service Date 7. Select Save * Expiration Date 31 • Linearity Lots screen appears with the new lot displayed Download to Instruments 7 Save Cancel 6 Effective Date: 12-Jul-2021 | HOOD05162003201651 Unrestricted © Siemens Healthineers, 2021 UniPOC™ Data Management System SIEMENS Healthineers Add Lot Information EDIT ANALYTE RANGES: Main Hospi EDIT LINEARITY LOT: Main Hospital INPUT INFORMATION Enter Linearity Ranges INPUT INFORMATION Reagent Type CHEM8+ Reagent Type CHEM8+ Lot No. 2287411090 1. In the Linearity Lot table, select desired linearity lot by selecting the checkbox in the Analyte * Lot No. 2287411090 Manufacturer column Manufacturer In-Service Date 31 2. Select Edit * Expiration Date 07/01/2021 31 In-Service Date 31 • The Edit Analyte Ranges screen appears with * Expiration Date 07/01/2021 Analyte CI stored information 31 3. Edit the range fields (L1-L10) as applicable Download to Instruments L1 To 4. Select Save L2 To L3 Previous 1 of 1 To Next Note: Ranges must be entered per analyte to L4 To Save Cancel generate a Linearity report. 3 L5 TO 12 Records Found L6 TO 2 Edit L7 TO Analyte A L1 L2 L3 L4 L8 To All L9 TO L10 To AnGap BUN 1 Previous 1 of 1 Next Crea 4 Save Cancel 7 Effective Date: 12-Jul-2021 | HOOD05162003201651 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. UniPOC and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc., or its affiliates. All other trademarks are the property of their respective owners. © Siemens Healthcare GmbH 2021 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 8 siemens-healthineers.com Effective Date: 12-Jul-2021 | HOOD05162003201651 Unrestricted © Siemens Healthineers, 2021
- unipoc