Using syngo.via WebViewer: Web Browser Extension Job Aid
This job aid explains how to access syngo.via WebViewer via a web browser, how to search for patients and studies, which visualizing features are available, and how to start a virtual conference to share your images.
Using syngo.via WebViewer – Web Browser Extension Job Aid syngo.via WebViewer The syngo.via WebViewer 1 is the mobile extension of your syngo.via, enabling mobile reading of images with diagnostic quality graphics and collaboration within the hospital network or on the go 2 . The syngo.via WebViewer is an iOS application supporting iPad, iPhone and iPod as well as your favorite browser 3 . The syngo.via WebViewer is indicated for reviewing medical images from syngo.via to support interpretation in the field of radiology, nuclear medicine and cardiology. This device should be used only when there is no access to a workstation. 1 The application syngo.via WebViewer is not for diagnostic viewing/reading on mobile devices in the US. Please refer to your sales representative whether the product is available for your country. Diagnostic reading of images with a web browser requires a medical grade monitor. For iPhone and iPad, country specific laws may apply. Please refer to these laws before using for diagnostic reading/viewing. For Japan: Applications on iPhone / iPad / iPod are not a medical device in Japan. Use at your own risk. They are not intended to be used for diagnosis. 2 Prerequisites include: wireless connection to clinical network, meeting recommended minimum hardware requirements, and adherence to local data security regulations. 3 Apple®, iPhone®, iPad®, iPod Touch® are trademarks of Apple Inc., registered in the U.S. and other countries. It has been observed, that syngo.via WebViewer VA20B is not working properly with Microsoft Edge. SIEMENS Restricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Healthineers Job Aid Effective Date: 22-Feb-21 I HOOD05162003167893 Job Aid I Using syngo.via Webviewer – Web Browser Extension Effective Date: 22-Feb-21 I HOOD05162003167893 1. Login syngo.via WebViewer extension requires a supported web browser 4 . No additional plugin is required. Simply type the syngo.via WebViewer server address in your browser. You can bookmark the server address in your browser to easily call the syngo.via WebViewer. Enter your username and password and click the Login button. webviewer.s s-healthmeers.com SIEMENS Healthineers syngo.via ¥2020.1.11.203 WebViewer Welcome to the syngo. via WebViewer. This server is freely available for theoven flash below to bege demonstration purposes. Please click an the Login buttor Login Terms of Use Quick Guide 4 Following web browsers and versions (no additional plugin is required): Microsoft Edge 83 and higher, Microsoft Internet Explorer 11, Mozilla Firefox ESR 68 and higher, Google Chrome 83 and higher, Apple Safari 13 and higher. SIEMENS Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Healthineers Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 22-Feb-21 I HOOD05162003167893 Job Aid I Using syngo.via Webviewer – Web Browser Extension Effective Date: 22-Feb-21 I HOOD05162003167893 2. Patient Browser After you have logged on with a new session, the search area is displayed with the Study List. The search area enables you to change the current search criteria by patient name, patient ID, date of birth, study date or modality. The following datasets are supported: • X-Ray • AX/XA • CT • MR • PET • PET-CT Fusion • Secondary Capture Images • Encapsulated PDF syngovia Weblewer x : + 0 X webviewer.siemens-healthineers.com Healthineers User- user22 Patient Browser × Patient Name: Date of Birth: Muddyvvv Modality Q 15 atiant ID: Study Date: Mwpdf Study List Name Birth Date Modality Study Date Patient ID 4D Mitral Valve 00/01/2020 CT 1300 Annionymous AAA Stont CTA 08/01/2020 C 1300 Annonymous Aortic Dissection Cin ... 01/01/2020 CT,SR 01/01/2020, 12:00 AM KEINEID Bi Phaso Liver CAP Ci ... 07/01/2020 CT 1300 Annonymous Carotid COW CTA On .... 07/01/2020 CT 1300 Annonymous Carotid Gliomit lumar 09/01/2020 1300 Annonymous ChestAbdPelvis 08/01/2020 CT 1300 Annonymous ChestAbdPelvis Cine ... 08/01/2020 CT 0100 Anonymous Coronary 101bpm Ci ... 07/01/2020 CT OTOD Annonymous COVID-19 AJAC,USB 01/01/2020 CT 2020-03-01 1200 COVID USB COMO-19_goUp199 .... 01/01/1997 CT,SFG,SR 2020-04-01 1200 goUp_199 COMID-19 goUp:99 .... 01/01/1997 CT 2020-04-01 1200 goUp_192 AIRC COMO 19 golip200 .- 04/01/2020 CT,SR, SEG 2020-04-01 1200 goup_200 CTA Renal Donor Cin .. 07/01/2020 CT 0100 Annonymous Luminos Agile Max, A .... 01/01/2015 CR 2015 01 01 0000 PID110 Mobilett Mira Max,C ... 01/01/2019 CR 2015-01-01,0000 Multitorn Hax, Chest ,... 01/01/198) CR, SC.SR 202D-01-01 0000 PIDSTO Multitom Rax, L-5pine .... 01/01/1975 CR.SC.5R 2020-01-01:0000 PID609 Multix Impact, Ankle ,... 01/01/2015 2020-01-01 :0000 PID420 PF 45ml Contrast 1300 Anonymous OS os OS SIEMENS Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Healthineers Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 22-Feb-21 I HOOD05162003167893 Job Aid I Using syngo.via Webviewer – Web Browser Extension Effective Date: 22-Feb-21 I HOOD05162003167893 3. Visualizing In the result list, double-click the desired study, enabling the images to be opened into the viewer. The syngo.via WebViewer provides access to image data with basic manipulation and measurement functions 5 as well as 3D postprocessing tools. Additionally, a comprehensive toolset, presets, layouts, and views are available. In addition, you can play 4D movies for appropriate data sets. The tools are in the Control area. SIEMENS Healthineers 2 Carotid Ctomie Tumor 1 2020.08-01 x User: useri 0#1063 597. CF Layout Clip/Slab A 5 Tools are provided for temporary image-manipulation only and cannot be saved. SIEMENS Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Healthineers Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 22-Feb-21 I HOOD05162003167893 Job Aid I Using syngo.via Webviewer – Web Browser Extension Effective Date: 22-Feb-21 I HOOD05162003167893 4. Printing In the Control area, select the Print the Image icon. You can adjust the printing settings and select the Print button. SIEMENS Carotid Cllomie Tumor X Healthineers User: user1 #1063 597, CT Layout Clip/Slab 15 LHE HQ A F A Layout Print 1 sheet of paper General Layt itst Destination md1 5b2ap Pages All Copies Layout Landscap Color Black and white More settings Clip/Slab HQ Print Cancel SIEMENS Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Healthineers Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 22-Feb-21 I HOOD05162003167893 Job Aid I Using syngo.via Webviewer – Web Browser Extension Effective Date: 22-Feb-21 I HOOD05162003167893 5. Conference 6 In some cases, you might need a second opinion or assistance when evaluating images with syngo.via WebViewer. The syngo.via WebViewer is also the tool to enable easy contact with referring physicians, colleagues on call, or clinicians working from home while providing treatment. The syngo.via WebViewer user can therefore easily discuss and share images on-screen. Start a virtual conference from anywhere on the syngo.via WebViewer and simply invite anyone for a shared case-review, a tumor board, or a seamless collaboration with fellow physicians 7 . All usual imaging functionalities are available to you, including our new Shared Visible Pointer 5 . Also included are a do-not-disturb feature for uninterrupted conversations and a simple and secure collaboration with affiliated hospital sites or colleagues on call. Conference Chip/sia's 5 Tools are provided for temporary image-manipulation only and cannot be saved. 6 Including on VA20 Version. Anyone participating in a conference needs to be registered as a syngo.via user. It is at our customers’ discretion whom they would like to give access to the syngo.via WebViewer and syngo.via. 7 The performance of the syngo.via WebViewer client depends on your network connection speed, the network latency between your location and the server as well as on the number of active users. In locations farther from for example the server, the performance can be slow. If you encounter a pop-up message stating “Currently no user license session available”, please wait a few minutes and try again. We recommend a wireless internet connection with a minimal speed of 3.5 Mbits. SIEMENS Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Healthineers Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 22-Feb-21 I HOOD05162003167893 Job Aid I Using syngo.via Webviewer – Web Browser Extension Effective Date: 22-Feb-21 I HOOD05162003167893 Conferencing x Received Invitations Sent Invitations Create Invitation Users Contact Information You can use this page to create an invitation for a collaborative conference. Fill in the information below and touch the `Create Invitation' button. Users added to invitation: Enter user to add to invitation: Subject Invitation Demo: 4:33 PM X This is a test invitation Comments Duration in Hours X Create Invitation Clear Do Not Disturb SIEMENS Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Healthineers Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 22-Feb-21 I HOOD05162003167893 Job Aid I Using syngo.via Webviewer – Web Browser Extension Effective Date: 22-Feb-21 I HOOD05162003167893 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2021 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com SIEMENS Healthineers
- snygo
- webviewer
- web-viewer
- web viewer