Viewing Previous PET QC Reports and Sinograms
Viewing Previous PET QC Reports and Sinograms All Biograph Systems Description: This step by step procedure will assist the user in viewing previous PET QC reports and sinograms. 1. From the Examination task card menu, select Setup à PET Quality Check 2. The PET Quality Check dialog appears. Go to the Configure tab. 3. Click View at the bottom of the dialog. PET Quality Check Perform and Monitor Configure Phantom Volume 6315 (cc) Isotope Ge-68 Assay Activity 2.44 mCi Assay Time 12:00:00 Assay Date 11-Jan-2011 Cross Calibration 1 M Partial Setup Time Alignment Full Setup Position phantom with laser Position phantom with last saved value QC Schedule day: Mon Tue Wed Thu Fr Sat Sun QC Schedule time: 06:30:30 Start View Cancel Help SIEMENS siemens.com/healthineers Healthineers Viewing Previous PET QC Reports and Sinograms Effective Date: 02/05/2018 | Quality HOOD05162002759036 4. A list of previously stored PET QC reports and sinograms are displayed. Click the report of interest and click Load. QC Results X Sinograms QC Reports PETQCSinogram_30Jul2009_065648.s PETQCSinogram_29Jul2009_081915.s PETQCReport_30Jul2009_065648 xml PETQCSinogram_29Jul2009 074853.s PETQCReport_29Jul2009_081915 xml PETQC Sinogram 27Jul2009 082122 s PETQCReport_29 Jul2009 074853 xml PETQCSinogram_27Jul2009_074314.s PETOCReport 27 Jul2009 082122 xml PETQCSinogram_26Jul2009_083742.s PETQCReport_27 Jul2009_074314 xml PETQCSinogram_25Jul2009_102752.s PETQCReport_26Jul2009_083742 xml PETQCSinogram_24Jul2009_074836.s PETQCReport_25Jul2009_102752 xml PETQCSinogram_23Jul2009_165150.s PETQCReport_24 Jul2009_074836 xml PETQCSinogram_23Jul2009_073341.5 PETQCReport_23Jul2009_165150 xml PETQCSinogram_22Jul2009_072414.s PETQCReport_23 Jul2009_073341 xml PETQCSinogram_21Jul2009_075734.s PETQCReport_22Jul2009_072414 xml PETQCSinogram_20Jul2009_083755.s PETQCReport_21Jul2009_075734 xml PETQCSinogram 17 Jul2009 084418.s PETQCReport_20Jul2009_083755 xml PETQCSinogram_16Jul2009_101725.s PETQCReport_17 Jul2009_084418 xml PETQCReport_16Jul2009_101725xml PETQCReport_15Jul2009_100637 xml Load Cancel 5. The PET Quality Check dialog appears. To view the report, click the View/Print Report button. PET Quality Check Perform and Monitor QC Configure QC Moving the bed to PET-FOV ... The bed is at PET-FOV. Gantry Check is being performed .. Gantry Check completed successfully. Partial Setup is being performed. Partial Setup completed successfully. Setting energy thresholds on ACS .. Energy thresholds set successfully. Configuring ACS ... Configured ACS successfully. PET Acquisition is in progress ... PET Acquisition completed successfully PET Image Reconstruction in progress ... PET Image Reconstruction completed successfully Report PASS Computed ECF 2.887e+007 View Sinogram View Trend View/Print Report Save&Close Post process Start Load Cancel Help Siemens Healthineers Services Page 2 of 4 Viewing Previous PET QC Reports and Sinograms Effective Date: 02/05/2018 | Quality HOOD05162002759036 6. If the sinogram is available, click View Sinogram. PET Quality Check Perform and Monitor QC Configure QC Moving the bed to PET-FOV ... The bed is at PET-FOV. Gantry Check is being performed .. Gantry Check completed successfully. Partial Setup is being performed. Partial Setup completed successfully. Setting energy thresholds on ACS ... Energy thresholds set successfully. Configuring ACS. Configured ACS successfully. PET Acquisition is in progress ... PET Acquisition completed successfully. PET Image Reconstruction in progress. PET Image Reconstruction completed successfully Report Computed ECF PASS 2.887e+007 View Sinogram View Trend View/Print Report Save&Close Post process Star Load Cancel Help 7. The Sinogram Viewer window will open. PET Quality Check Perform and Monitor OC Coffigure QC Setting energy thresholds on ACS Energy thresholds set successfuly Configuring ACS. Sinogram Viewer X DEA Block 2009/02/06 13:36:28 44. 45 Start Cancel Help Siemens Healthineers Services Page 3 of 4 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instruc- tions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety infor- mation like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018
- Viewing Previous PET QC Reports and Sinograms