Syva® Viva-ProE™ System As Needed Maintenance Job Aid

This job aid provides instructions on how to perform as-needed maintenance on the Syva® Viva-ProE™ System.

As-Needed Maintenance Replace the Cuvette Rotor From the Home screen select the Tasks icon 1. Select the Maintenance tab 2. Remove the Cuvette Rotor Cover 3. In the software, select Replace Cuvette Rotor 4. Pull up on the Stirrer unit using the Stirrer Knob [1] O 5. Remove the cuvette rotor by pulling up using the Center Knob [2] 6. Install the new cuvette rotor by carefully aligning the O four notches of the new cuvette above the slits on the rotor holder (the cuvette rotor will drop down) 7. Push down the Stirrer Knob 2 8. In the software, select Continue 210 9. Select Yes at the “Reset Cuvette Rotor Counter” prompt 10.Select the Reset icon to reset the system and return to a stand-by state 11.Replace the cuvette rotor cover 12.Perform a cuvette rotor blank  Home screen > Start of Day > Check Cuvette Blank  Select Run > Yes Note: When the cuvette blank is within the defined acceptable ranges, the measurement is automatically accepted and the information on the screen is updated. No further action is required. HILS#705 Effective date 06/01/2016 © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. As Needed © Siemens AG 2013. All rights reserved. Maintenance As-Needed Maintenance Fill the Cooling Fluid 1. Remove the Front Panel from the instrument by lifting it up from the bottom and pulling it forward. 2. Check the Cooling Fluid level in the reservoir. The level should be between the minimum [3] and maximum [2] markers. 2 Note: If the cooling fluid level is below the Minimum [3] marker, proceed to step 3. 3. Fill the Cooling Fluid reservoir: 3  Remove the cooling fluid reservoir cap [1]  Add Cooling Fluid until the level is just below the maximum marker [2]  Reattach the cap [1]  Re-install the front instrument panel Note: The Cooling Fluid must be diluted with distilled water. Use 1 part Cooling Fluid to 4 parts distilled water. © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. © Siemens AG 2013. All rights reserved. As-Needed Maintenance Clean Reagent Rotor Compartment 1. Cap the reagent bottles that are on the Reagent Rotor 2. Remove the Reagent Rotor with the capped reagent bottles and store appropriately following standard laboratory practices 3. Remove the Reagent Rotor from the system:  Remove the Reagent Rotor cover  Remove the knob from the center of the rotor  Lift the Reagent Rotor out of the instrument 4. Clean the Reagent Rotor:  Pour all condensation fluid out of the rotor  Dry the inside and outside of the rotor with a paper tissue or a clean, dry cloth Dry the Stator (base on which the rotor rests) with 13 5. a paper tissue or a dry, clean cloth 6. Reinstall the Reagent Rotor:  Align the white dot on the rotor with the dot on the driving axle  Reinstall the knob 7. Uncap the reagent bottles 8. Replace the Reagent Rotor cover © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. © Siemens AG 2013. All rights reserved.

  • Pro E
  • Pro-E
  • VivaProE
  • VivaPro E
  • Viva ProE
  • VivaPro-E
  • Viva E
  • VivaE
  • Viva-E
  • DOA
  • DOA Testing
  • Drugs of Abuse Testing
  • replace cuvette rotor
  • cuvette rotor
  • counter
  • check cuvette
  • reset cuvette rotor counter
  • stirrer knob
  • reagent rotor
  • cooling fluid