Syva® Viva-ProE™ System Daily Maintenance Job Aid
This job aid provides instructions on how to perform daily maintenance on the Syva® Viva-ProE™ System.
Start of Day / Daily Maintenance On the Home screen select Start of Day Start of Day 1. Empty diluted and concentrated waste containers following standard laboratory operating procedures 2. Fill treated water container Start of Day Check for sufficient amount of water in treated water container to run samples. If needed, fill an empty 5 L water Description Status container with distilled or deionized water. Add 12.5 mL of 1 Empty Waste System Solution 3. Check syringes during Prime System 2 Fill treated water container From the Home screen, select Tasks > Maintenance > 3 Prime and Drain > Prime System Check syringes Verify no air bubbles or leaks are present 4 Check cuvette blank Remove caps off reagent bottles in the reagent rotor Check/refill 10% cleaning solution 4. Check cuvette blank If the cuvette blank results are within the defined acceptable 6 Check daily report ranges, the measurement is automatically accepted and the information on the screen updated. No further action is 7 Check 0. 1N HCI Cleaning Solution required. 8 Check Reagent Inventory If the cuvette blank results are not within the defined 9 acceptable ranges, refer to the Troubleshooting section of Remove Caps this guide for details 5. Check / refill 10% cleaning solution (Probe rinse) Fill the tube located in the wash on the sample rotor with Note: The minimum Start of Day tasks include 1 cleaning solution through 6 shown above. This checklist can be edited when logged into the system as an Administrator. If the cleaning solution has expired, select Check/refill 10% cleaning solution, then select Replace > enter Lot Number, HILS#705 Effective date 06/01/2016 Expiration Date > Accept © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. Start of Day / Daily © Siemens AG 2013. All rights reserved. Maintenance Start of Day / Daily Maintenance Reagents admin 4/29/2014 8:46:02 AM Print Help 6. Check daily report Filter by Test Clear 8 Reagent Info Select tabs to view expired or expiring Pos Tests Reagent Status Load Standar R4 T50 Unload Switch Rotor calibrator/control lots, reagents and calibrator and R5 CR R6 6 AM Replace 7 2 2 2 control planning R7 R3 R8 Select maintenance planning to view tasks that are R9 R10 required R11 R17 R2 R12 Probe rinse Select Daily Report to print a copy of the report R13 Acid Solution 7 R14 7. Check 0.1N HCl Cleaning Solution (Acid Solution) R15 R16 R17 Fill the bottle located in the reagent rotor C15 R2 R18 OP3 R2 R19 T50 R2 If the cleaning solution has expired, from the R20 CR R2 R13 R21 BAM R12 Reagents screen, select Replace, enter Lot Number, R22 Expiration Date, and select Accept QC 8. Check Reagent Inventory Stop Home Reagents Calibration Central Patients Sean Results Unload OC Events Select the Reagents icon Reagents admin ? Select Reagent Info 4/29/2014 8:43:50 AM Logout Print Help Filter by Test Clear If reagent is flagged (see Status Icons), the bottle 8 Print Hide Info POS Tests Reagent Status Lot Number Lot Expiration requires refill 2 5 Onboard Stability Remaining Volume Remaining Tests Completed Tests CR 7/31/2014 10.9 mL 38 R6 6AM R1 F4A 11/30/2014 9d 12.5 ml 68 12 Under Remaining Tests column, view test number R7 RB R9 if less than 10, fill the bottle R10 – On the Reagents screen, select the bottle to refill, R11 R12 Probe rins 1007 7/23/2014 28.0 ml 16.9 mL select Replace, enter Lot Number, Expiration Date, R13 Acid Solution 45158 5/3/2015 R14 R15 and select Accept R16 AMS 7/31/2014 10.3 ml R17 C15 R2 F4A 8/30/2014 10 8.7 ml 103 Note: If the lot # of the reagent has changed, R18 OP 3 R2 F3 11/30/2014 5h 3.8 mL 47 31 R19 T50 R2 E8 7/31/2014 5h 4.7 ml 54 calibration is required to be performed R20 CR R2 F1 GAM FAA 7/31/2014 5h 5.6 mL 97 R21 11/30/2014 0.0 ml 9. Remove Caps off reagent bottles R22 R23 R24 Qc Stop Home Reagents Calibration Central Patients Scan Results Unload OC Events © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. © Siemens AG 2013. All rights reserved. Start of Day / Daily Maintenance Request and Run Controls From the Home screen select Control Control Load Controls admin 1 4/11/2014 8:22:46 AM Logout Print Help 1. Select Load for each control Control Tests Control Lot Pos 37 Level 0 (Negative) AM5 6AM C15 OP3 T50 A1 Unload 2. Load the controls into their assigned positions Level 5 (Positive) AM5 C15 OP3 T50 F5 AZ Level 4 (6-AM) 6AM F A3 A10 on the sample rotor UTAK Validity 3 CR A1035 AA 32 A11 UTAK Validity 4 CR A1036 A5 3. Lower glass lid, select Process > Yes 49 A12 Process A7 28 W 27 26 4. If internal barcode scanner is on, select Scan 2 25 prior to process A6 23 Scan 20 21 Note: Level 0 (Negative) and Level 5 (Positive) 19 12 controls are run daily to validate the existing qualitative DAT calibration curve. 3 Review UTAK controls are run daily to validate the validity QC calibration curve. Stop Home Reagents Calibration Control Patients Results Unload ac Events © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. © Siemens AG 2013. All rights reserved. Start of Day / Daily Maintenance Evaluating Qualitative DAT & Validity Controls 1. On the Results screen select the Calibrators Result and Controls radio button 1 admin 4/2/2014 10:53:59 AM Logout Print Help Reagent Blanks Calibration and Controls Patients Search 2. Select the test to display the measurement results Test Amphetamines 500 - Emit Cal Level 2 Amphetamines 500 - Control 3. Verify the Level 0 result as negative and Level 5 result 4 AM5 2 Level Value dAbs/m Flags Control Result Range as positive 6AM Level 2 500.0000 0.3961 Level 0 3 (Negative) Negative 0.34580 4. Results automatically accept if there are no flags C15 Level 5 (Positive) Positive 0:54211 V except for v and V OP3 lowercase v indicates negative T50 Details Details uppercase V indicates positive CR 0.3961 5. If the result is flagged, select from the following 0.3169 0.2377 options: 0.1585 Select Repeat to rerun the control measurement 0.0792 .. ...... 0.0000 Absorbance (dAbs/min) Select Accept to accept the control measurement 50 0 50 100 150 200 250 300 350 400 450 500 550 Concentration result A Status: Accepted 4/2/2014 10:27 AM Select Reject to reject the control measurement Export Print Daily QC Accept -Curve Reject Curve result O QC Note: If Repeat is selected, the Viva-ProE™ system will Stop Home Reagents Calibration Control Patients Resume Results Unload Qc Events average the results of both replicates. Siemens recommends selecting Reject, and then rerun the control Note: The system will not run test(s) requests in which the control(s) are flagged unacceptable or out of range. © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. © Siemens AG 2013. All rights reserved.
- Pro E
- Pro-E
- VivaProE
- VivaPro E
- Viva ProE
- VivaPro-E
- Viva E
- VivaE
- Viva-E
- DOA
- DOA Testing
- Drugs of Abuse Testing
- check waste container
- empty waste container
- fill treated water
- prime and drain
- prime system
- probe rinse
- daily report
- acid solution
- reagent inventory
- reagent
- run controls
- calibration
- calibrator