Syva® Viva-ProE™ System Monthly Maintenance Job Aid
This job aid provides instructions on how to perform monthly maintenance on the Syva® Viva-ProE™ System.
Monthly Maintenance Clean Waste and Treated Water Containers From the Home screen select the Tasks icon 1. Select Maintenance > Clean Waste and Treated Water Container 2. Empty the treated water container ED 3. Pour 0.1N NaOH (Sodium Hydroxide) to coat the bottom of the Treated Water container DILUTED WASTE 4. Shake the container to distribute the NaOH 5. Pour the NaOH from the Treated Water container into the empty Diluted Waste container and repeat steps 3 and 4 6. Pour the contents into the empty Concentrated Waste container and repeat steps 3 and 4 7. Rinse the Treated Water container only with distilled water three times to remove the NaOH 8. In the Maintenance screen, select Confirm 9. Select Reset > Yes to reset the system and return to a stand-by state (1) Rese HILS#705 Effective date 06/01/2016 © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. Monthly Maintenance