Syva® Viva-ProE™ System Semi-Annual Maintenance Job Aid
This job aid provides instructions on how to perform semi-annual maintenance on the Syva® Viva-ProE™ System.
Semi-Annual Maintenance Replace the Stirrer Belt From the Home screen select the Tasks icon 1 1. Select Maintenance > Replace Stirrer Belt 2 2. Remove cuvette rotor cover W 3. Remove stirrer belt [2] from the pulley [1] HAA 0 210 0 4. Starting from the eccentric pulley [3], place the new stirrer belt around the pulley 5. Repeat steps [3] and [4] for the second pulley 6. In the software, select Confirm 7. Select the Reset icon to reset the system and return to a stand-by state Note: A pair of tweezers is useful when performing this task HILS#705 Effective date 06/01/2016 © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. Semi-Annual Maintenance Semi-Annual Maintenance Replace the Water Filter From the Home screen select the Tasks icon 1. Select Maintenance > Replace Water Filter 2. Remove the Treated Water Container from inside the cabinet 3. Remove the Treated Water Container cap [1] 4. At the end of the tubing [3], remove the water filter [5] by unscrewing from the filter holder [4] Note: To avoid contamination of the water, do not lay the tubing with the filter on an unclean surface W TREATED WATER 5. Replace with a new filter by screwing back on the filter holder 6. Place the tubing with filter back inside of the Water Container and secure the cap 7. Place the Water Container back inside the cabinet 5 8. In the software, select Confirm 9. Select the Reset icon to reset the system and return to a stand-by state © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. © Siemens AG 2013. All rights reserved. Semi-Annual Maintenance cap Clean the System From the Home screen select the Tasks icon 1. Select Maintenance > Clean System 2. Empty the Diluted Waste container 3. Remove the cap and tubing with the attached filter from the Treated Water container tubing 4. In the software, select Clean System – the system will cycle for approximately 30 minutes while it flushes TREATED all of the fluid from the instrument Note: Operator assistance is required during this entire procedure. The system software will prompt the operator to perform a task and wait for confirmation after each step. This procedure can take up to 4 hours to complete. 5. Fill a container with approximately 500 mL of Sodium Hypochlorite System Cleaning Solution. Be sure to use a container that will allow the filter from the Treated Water tubing to be completely submerged in the system cleaning solution. 6. When prompted, insert the Treated Water tubing into the Sodium Hypochlorite System Cleaning Solution container 7. In the software, select Continue – the system will cycle for approximately 15 minutes 8. When prompted, remove the tubing out of the Sodium Hypochlorite System Cleaning Solution container 9. Select Continue – the system will cycle for approximately 30 minutes while it flushes the instrument 10.Clean the Treated Water container (refer to Monthly Maintenance activities) and refill with freshly prepared treated water 11.When prompted, insert the tubing into the filled Treated Water container 12.Select Continue – the system will cycle for approximately 15 minutes 13.When prompted, remove the tubing from the Treated Water container 14.Select Continue – the system will cycle for approximately 30 minutes Note: The system will prompt the Operator to repeat steps 11 through 14 three times. Once the process has completed, the system will automatically return to a stand-by state. GPE#06005.008 Effective date XX/XX/2016 Perform a Prime System and check the syringes for air bubbles. © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. © Siemens AG 2013. All rights reserved.
- Pro E
- Pro-E
- VivaProE
- VivaPro E
- Viva ProE
- VivaPro-E
- Viva E
- VivaE
- Viva-E
- DOA
- DOA Testing
- Drugs of Abuse Testing
- stirrer belt
- water filter
- clean system