Cleaning the Ancillary Probe Video
This video provides instructions on how to properly clean the ancillary probe on the ADVIA Centaur® XP system.
After the mechanics have been turned off, lift the instrument lid. Move the in process Q to the right. Pull the probe tordue by holding the top of the probe assembly. Use a gentle downward motion to wipe each probe with lint, free tissue or gauze saturated with cleaning solution. Thoroughly rinsed the reagent probe with lint free tissues or gauze saturated with DI water. Close the instrument lid and turn system mechanics on.