PEPconnect

Cleaning the Reagent Probes Video

This video provides instructions on how to properly clean the reagent probes on the ADVIA Centaur® XP system.

After the mechanics have been turned off, lift the instrument and reagent probes covers. Push the reagent probe assemblies tored the incubation ring by pushing the Motors. Use a gentle downward motion to wipe each probe with lint, free tissue or gauze saturated with cleaning solution. Thoroughly rinsed the reagent probe with lint free tissues or gauze saturated with DI water. Close the instrument lid and turn system mechanics on.

  • monthly maintenance
  • centauro
  • centuar
  • centuarxp
  • centaurxp
  • centauroxp
  • centuarxp
  • cenatur
  • cenaturxp
  • cen
  • cenxp
  • centar
  • centarxp
  • centau
  • centauer
  • centauerxp
  • centure
  • centurexp