Performing Maintenance - Reinstalling Job Aid
This job aid provides instructions on how to perform Maintenance on your CLINITEK Advantus® Analyzer, including activities for reinstalling the parts.
Maintenance - Reinstalling Reinstalling the Moving Table Reinstalling the Holddown Plate 1. Hold the table with the small rectangular tab 1. Position the holddown plate with the arrow side facing to the back facing up and the arrow pointing towards the 2. Align the 2 grooves on the bottom of the table back with the edges of the platform on which the table 2. Place the pin on the front of the holddown plate rests into the hold at the front of the fixed platform 3. Gently push the table in until you hear the tab 3. Align the tab at the back of the holddown plate latch into the hold position with the slot at the back of the platform 4. Check that the table is secure 4. Snap the holddown plate into place. Listen for a loud click, indicating proper installation 5. Ensure the white calibration bars are visible GPE04001.005 Effective Date: 09/12/2014 Maintenance - Reinstalling Reinstalling the Fixed Platform Reinstalling the Push Bar 1. Align the 2 grooves on the bottom of the fixed 1. Hold the push bar at the indented end platform with the arms extending forward from the analyzer 2. With the end slightly upward, insert the peg on • The flanges on the sides of the holddown plate the other end of the bar into the hold in the push align just outside the read area cover. The top edge mechanism of the platform aligns just under the cover 3. Lower the push bar into place 2. Gently push the platform in as far as possible • Push past the ridge to correctly position the Reinstalling the Waste Bin Liner platform 1. Place a new liner into the waste bin • Caution: Do not for the platform. You may damage the moving table or fixed platform if incorrectly positioned GPE04001.005 Effective Date: 09/12/2014
- Maintenance
- Reinstalling
- Test Table insert
- routine
- periodic
- disinfecting
- calibration bar