CLINITEK Novus™ Automated Urine Chemistry Analyzer Quality Control Online Training
State the materials required to prepare controls. List the steps necessary to perform a quality control test. Locate and review quality control results. Identify and resolve common quality control errors. This clinical laboratory training qualifies for continuing education units (CEU).
Welcome to the CLINITEK Novus™ Automated Urine Chemistry Analyzer Quality Control Online Training course. This course includes information regarding how to run controls and review the results. Select Next to continue. Identify and resolve common quality control errors List the steps necessary to perform a quality control test Locate and review quality control results State the materials required to prepare controls Upon successful completion of this course, you will be able to: Select Next to continue. Congratulations. You have completed the CLINITEK™ Novus Quality Control Online Training course. Listed below are the key points that have been presented. Take time to review the material before you proceed to the final quiz. State the materials required to prepare controls Materials required: Sample tubes Sample rack Distilled or deionized water List the steps necessary to perform a quality control test Prepare control material Pour at least 2mL of control material into tubes and place in sample rack Load sample rack onto the system Select Control Enter the operator’s initials, if prompted, and select Enter. Follow the instructions on the screen. Select an Enable option for each control lot that you want to use. To change a lot number or add a lot number, select Change for the control that you enabled. Follow the instructions on the screen. After placing the rack of controls on the rack handler select Start to begin the run. Locate and review quality control results At the end of the control run, results are located on the Results Details screen. Review each result to determine if the control results are acceptable. To review results at the end of a QC run: If your instrument utilizes the Control Check feature to check the upper and lower reporting limits for controls, the CLINITEK Novus™ system will flag any results that are out of the expected range. Note: If the Control Check feature is disabled, the instrument will not flag abnormal results. If your instrument is not setup to flag results that are out of the expected range, utilize the expected control values from the QC package insert to determine whether the control results are acceptable. To recall the Control Results Details screen at another time, from the Home screen, select Results > Control and select the search criteria. Identify and resolve common quality control errors The control results should match the values in the control package insert. Controls need to be prepared correctly and cannot be expired. Controls are expired: Prepare new controls if stability is exceeded Open a new lot of control if strips are expired System errors: Tube Error: Remove caps from tubes Low Volume: Inadequate sample volume Select Next to continue. Bienvenue à la Formation en ligne de l'analyseur de gaz du sang RAPIDPoint® 500, Etalonnage. Les étalonnages du système en routine sont nécessaires pour maintenir la précision des résultats de l'analyseur et ceci est valable pour tout analyseur électrochimique. Tous les réactifs d'étalonnage sont contenus dans la cartouche de mesure, qui est remplacée chaque mois ou lorsque les tests disponibles sont consommés. The system can run both Siemens and non-Siemens control material. When used with the CLINITEK Novus™ Automated Urine Chemistry Analyzer and CLINITEK Novus™ 10 Reagent Cassette Pak, CLINITEK Atlas® Positive and Negative Control Strips provide defined results for: color clarity glucose bilirubin ketone (acetoacetic acid) specific gravity blood pH protein urobilinogen nitrite leukocyte esterase Note: When using the CLINITEK Novus™ PRO 12 cassettes, controls that contain creatinine and albumin at the microalbuminuria level must be purchased from another vendor since these parameters are not included in CLINITEK Atlas® negative and positive controls. Siemens Quality Control Preparation Learn about Siemens quality control preparation. Slide NumberText BlocksCalloutsAudio ScriptImage File1 Materials required: Sample tubes CLINITEK Atlas® Control Strips Sample rack Select Next to continue. Note: If audio does not automatically start, select the play arrow in the top left to begin.Sample tubes, CLINITEK Atlas Control Strips, and a Sample rack are needed to prepare the control material.2The control solutions are stable for 8 hours after preparation, except for bilirubin in the positive control, which is stable for 3 hours under normal laboratory conditions. Select Next to continue.With the exception of bilirubin in the positive control, the CLINITEK Atlas control solutions are stable for 8 hours after preparation under normal laboratory conditions. Bilirubin is stable for 3 hours.3Place 12 mL of distilled or deionized water into appropriately labeled specimen tubes.Select Next to continue.Put 12 mL of distilled or deionized water into appropriately labeled specimen tubes. Do not use tap water.4Remove a positive and a negative control strip from the bottle and replace the cap immediately.Place each strip into the appropriate tube. Cap the tubes. Select Next to continue.Remove the control strips from the bottles and replace the caps immediately. Make sure that the caps are tight. Place the Positive and Negative strip into the appropriate tube. Cap the tubes.5Gently invert the tubes for 2 minutes.Allow the tubes to stand for 30 minutes at room temperature.Select Next to continue.Gently invert the tube(s) for 2 minutes then, allow them to stand at room temperature for 30 minutes. 6Invert one more time, then remove and discard the strips. Select Next to continue.Invert one more time, then remove and discard the strips according to your standard laboratory procedures.7 When using lipless tubes, place the tubes in a rack with the first control in position one and the second in position two. Lipped tubes need to be placed in every other tube slot in the rack. Minimum volume for running any control is 2 mL. When complete, select the X in the upper-right corner to close the window and continue.Quality control tubes are placed in a rack with a tube in position one and another in position two. If you use lipped tubes (for example, Urin-Tek tubes), place them in every other tube slot in the rack. Minimum volume for running any control is two milliliters. Non-Siemens Quality Control Preparation Learn about non-Siemens quality control preparation. Follow instructions for use to prepare non-Siemens controls. Add at least 2mL of quality control material into tubes. Quality control tubes are placed in a rack with a tube in position one and another in position two. If you use lipped tubes (for example, URIN-TEK® tubes), place them in every other tube slot in the rack. Minimum volume for running any control is two millimeters. When complete, select the X in the upper-right corner to close the window and continue. L'étalonnage complet se produit toutes les 8 heures et nécessite 5 minutes 30 secondes. Il comprend toutes les fonctions de l'étalonnage 1 et 2 points, plus une pente Cooximetre utilisant un étalon supplémentaire compris dans la cartouche de mesure. Control solutions should be used on a regular basis to verify proper functioning of the CLINITEK Novus™ Automated Analyzer and the cassette. Perform a quality control test after calibration according to your laboratory practices. Performing Quality Control Tests Learn how to perform a quality control test on the system. Instructions:If media does not automatically start, select the play arrow to begin.Flash File:/content/generator/Course_90006097/Sim_RunQC_9/1. Sim_CN_QC_RunQC_29Oct2012_9.swfHTML5 File:/content/generator/Course_90006097/Sim_RunQC_9/index.htmlPDF File: À l'issue de ce cours, vous serez en mesure d'identifier les éléments d'un étalonnage 1 point, d'un étalonnage 2 points, d'un étalonnage complet, de faire la distinction entre interruptible et non interruptible, d'identifier les conditions dans lesquelles l'étalonnage rétrocal est actif et consulter les données d'étalonnage. You have just learned about performing Quality Control Testing. Select the link below to practice what you have learned. Practice: Performing Quality Control Testing Practice performing quality control testing. Instructions:If media does not automatically start, select the play arrow to begin.Flash File:/content/generator/Course_90006097/sim_CN_QC_RunQCYourTurn_09Aug2012-9/sim_CN_QC_RunQCYourTurn_09Aug2012-9.swfHTML5 File:/content/generator/Course_90006097/sim_CN_QC_RunQCYourTurn_09Aug2012-9/index.htmlPDF File: La majorité des échantillons artériels sont recueillis au moyen d'un dispositif spécialisé. Cette méthode offre une héparinisation adéquate de l'échantillon en permettant l'anticoagulation et la prévention contre la formation de caillots. Après avoir vérifié que tout l'air a été retiré du dispositif, mélanger l'échantillon. Comme il est fréquent que les globules rouges sédimentent, même sur une courte période de temps, mélanger permet une uniformité de l'échantillon mesuré. Control results are located on the Control Results Details screen which automatically appears at the end of a QC run. To review results at the end of a QC run: If your instrument utilizes the Control Check feature to check the upper and lower reporting limits for controls, the CLINITEK Novus™ system will flag any results that are out of the expected range. Note: If the Control Check feature is disabled, the instrument will not flag abnormal results. If your instrument is not setup to flag results that are out of the expected range, utilize the expected control values from the QC package insert to determine whether the control results are acceptable. To recall the Control Results Details screen at another time, from the Home screen, select Results > Control and select the search criteria. Reviewing QC Results without Control Checks Learn how to review QC when your instrument is not configured to flag abnormal results. Instructions:If media does not automatically start, select the play arrow to begin.Flash File:/content/generator/Course_90006097/Sim_CN_QC_ReviewResults_800x477_9/Sim_CN_QC_ReviewResults_800x477_9.swfHTML5 File:/content/generator/Course_90006097/Sim_CN_QC_ReviewResults_800x477_9/index.htmlPDF File: Reviewing QC Results at the End of a QC Run Learn about reviewing quality control test results at the end of a QC run. Checklist TitleChecklist TypeChecklist ContentReview DetailsHTML At the end of the control run, the Control Results Details screen automatically displays. Use the scroll button to navigate the details at the top of the screen. Select each checkbox to learn more about reviewing quality control results at the end of a QC run. Review ResultsHTML Review the results in the analytes area. Any result that is out of the expected range will be flagged with an asterisk (*). The system will not process any patient samples until controls are reprocessed and all results are in the expected range. Use the navigation buttons at the bottom of the screen to view other levels of control. Print or Send ResultsHTML To print results or send results to the LIS (Laboratory Information System), select a function button on the right side of the screen. Return to the Home ScreenHTML After the QC review is complete, return to the Home screen by selecting the back arrow and then the Home button. When complete, select the X in the upper-right corner to close the window and continue. Recalling Quality Control Results Learn about recalling quality control results. Checklist TitleChecklist TypeChecklist ContentSelect ResultsHTML Control results can be accessed at any time. Select Results from the Home screen. Select each checkbox to learn more about recalling quality control results. Select ControlHTML Select Control. Select Search CriteriaHTML Select the search criteria: Date Range, Control Lot, Cassette Lot, Operator ID, All Results, Sequence Number. Review ResultsHTML The Results Control screen displays the results that meet the search criteria. The most recent results display at the top of the control results summary. To navigate the summary, select a scroll arrow. To view, delete all results, print all results, send all results to the LIS or graph results, select a function button on the right side of the screen. View Control Result DetailsHTML Select a result in the control results summary and select View. Review DetailsHTML To navigate the details, use the scroll arrows. Review the results in the analyte area. Any result that is out of the expected range will be flagged with an asterisk (*). The system will not process any patient samples until controls are reprocessed and all results are in the expected range. Use the navigation buttons at the bottom of the screen to view other levels of control. Print or Send ResultsHTML To print results or send results to the LIS (Laboratory Information System), select a function button on the right side of the screen. Return to the Home ScreenHTML After the QC review is complete, return to the Home screen by selecting the back arrow and then the Home button. When complete, select the X in the upper-right corner to close the window and continue. The control results should match the values in the control package insert. Controls need to be prepared correctly and cannot be expired. Common QC Errors Learn about common QC errors. Tab TitleTextControl Ranges Exceeded Controls are expired: Prepare new controls if stability is exceeded Open a new lot of control if strips are expired System Errors Tube Error: Remove caps from tubes Low Volume: Inadequate sample volume When complete, select the X in the upper-right corner to close the window and continue.
- clini
- novis
- qc
- clinical laboratory training