
CLINITEK Status® Portfolio for Point of Care Maintenance Online Training
Clean instrument interior and exterior. Clean the calibration bar, test table and test table insert. Disinfectithe calibration bar.Change the batteries.
Identify the steps to perform As Needed Maintenance Tasks Identify the steps to perform Daily Maintenance Tasks Welcome to the CLINITEK Status®+ Analyzer Maintenance online module. After completing this course, you will be able to: Select Next to continue. Key Maintenance Tasks: Cleaning the instrument Cleaning the test table and insert Disinfecting the test table and insert Cleaning the calibration bar Changing batteries Maintenance Tasks Learn how to perform each maintenance task. Checklist TitleChecklist TypeChecklist ContentCleaning the instrumentHTMLSelect each checkbox to learn how to perform Daily and As Needed maintenance tasks on the CLINITEK Status®+ Analyzer. Keep the outside of the CLINITEK Status®+ Analyzer clean and free of dust. Daily, turn the analyzer off and then wipe the outside, including the display area, with a damp cloth and a mild detergent.Cleaning the test table and insertHTML For routine cleaning, remove the insert. Clean thoroughly and then rinse both sides under running water. Dry and replace the insert. As needed, clean the test table. Pull the test table slowly out of the analyzer while the analyzer is powered on, and remove the insert. Drain the drip tray if necessary. Clean the test tables with a cotton-tipped swab moistened with water and then dry thoroughly with a soft cloth. (Do not clean the Calibration Bar). Reinsert the test table and insert.Disinfecting the test table and insertHTML Place the insert and test table into a disinfecting solution with the white calibration bar above the liquid level. Allow the test table and insert to soak between 2 to 10 minutes. Rinse the test table and insert thoroughly with water and then dry with a soft cloth. Replace the test table and insert into the analyzer.Cleaning the calibration barHTML Remove the insert from the test table and then remove the test table from the analyzer. Check the white calibration bar for dirt or discoloration. Gently clean the area with a cotton-tipped swab or soft cloth moistened with distilled water. After the calibration bar has air-dried, inspect the surface. If the calibration bar cannot be cleaned or is still marked, replace it with a new test table. Reinsert the test table and insert into the analyzer after cleaning is completed.Changing batteriesHTML The battery icon appears in the upper-right corner of the title bar when the analyzer is being powered by batteries. The number of segments displayed represents the amount of power remaining in the batteries. The analyzer is designed to process approximately 100 tests from a set of batteries. The Power Save feature is always activated when the instrument is powered by batteries. If the instrument is not used in 5 minutes when battery-powered, it will automatically power down. With the test table and insert removed, place the analyzer on one side. Remove the battery cover by pressing down on the tab and pulling out. Remove the current batteries and replace with a new set of 6 AA-size batteries. Replace the battery cover and turn the instrument back onto its base. Replace the test table and insert into the instrument. Congratulations. You have completed the CLINITEK Status®+ Analyzer Maintenance module. Once you have reviewed the content on this page, you may proceed to the assessment portion of the course. In this course, you have learned how to: Clean the instrument Daily, keep the outside of the CLINITEK Status®+ Analyzer clean and free of dust. Clean the test table and insert Use a cotton-tipped swab moistened with water to routinely clean the test table and insert. Use care not to touch the white calibration bar. Use personal protective equipment. Follow the on-site procedure of your facility on frequency of cleaning the test table and insert. Disinfect the test table and insert Disinfecting the test table and insert is necessary if the analyzer will be transported to another location or if moved to another site which involves going through non-biohazard areas. Use deionized water and 5% sodium hypochlorite as a disinfecting solution. Soak the test table and insert in cleaning solution for a minimum of 2 minutes and maximum of 10 minutes. Ensure that cleaning solution does not come in contact with the white calibration bar. Clean the calibration bar Regularly check the white calibration bar on the test table, and always check it after a strip jam. If the white calibration bar is dirty or discolored, gently wipe and clean it with a new cotton-tipped swab or lint-free cloth moistened with distilled water. Use care not to scratch the white calibration bar. Solvents of any kind must not be used to clean the bar. Change the Batteries The battery power icon appears in the upper-right corner of the title bar when the analyzer is being powered by batteries. The number of segments displayed represents the amount of power remaining in the batteries. Replace with new AA-size batteries, as needed. Select Next to continue.
- clini
- connect
- +
- plus
- daily maintenance
- as needed maintenance
- clinical laboratory training