(D13) Dot Add-in Configurator Protocols_Editing GSP
(D13) Dot Add-in Configurator Protocols_Editing GSP
(D13) Dot Add-in Configurator Protocol – Editing GSP 1. Open > Exam Explorer > USER tree 2. Select > Region (Head) > Dot Engine > Program folder > DOT (Brain) 3. Select > Exam Strategy branch (e.g., t1_sag) 4. Select > Strategy (e.g., Standard) o Select Sequence > Right mouse > Select Properties 5. Select > Dot add-in tab > Edit configuration Exam Explorer - \USER\head Brain Dot Engine\program\t1_sag\Standard X Object Edit Tools Insert Help SIEMENS .................. 1 [1 f12d_sag 02:28 Generic Views USER head Brain Dot Engine y program Dot Brain +1 localizer mt1 sag ...... Standard Resolution fod ..... Speed focus Motion-insens ...... + ·15 t2 ......... + dark-fluid 1 epi_diffusion mit1 tra_pre + contrast agent clinical libraries + advanced applications programs library Protocol properties - Dot add-in - t1_f12d_sag X Standard Dot add-in Execution Auto load Copy References Dot add-in: Generic Views Edit Configuration ... Reset Configuration OK Cancel Help SIEMENS www.usa.siemens.com/healthcare 595 Effective Date: 09/06/2016 Page 1 of 3 6. Select > Setup button WUSER\head\Brain Dot Engine\program\t1_sag\Standard\t1_f12d_sag Details Dot views TA: 2:28 PM: REF PAT: Off Voxel size: 0.7x0.7x4.0 mm Rel. SNR: 1.00 fl Slice group 1 + FoV read 220 Slices 25 FoV phase 100.0 % Dist. factor 30 Slice thickness 4.0 Position Isocenter TR 440. ms Orientation Sagittal ms Phase enc. dir. A >> P TE 2.46 AutoAlign Head > Basis Averages 1 Phase oversampling 30 Concatenations 1 Filter |Prescan Normalize, E Coil elements |HE1-4 Routine Contrast Resolution |Geometry System Physio Inline Sequence OK Cancel Virtual Coils .. Setup .. Help 7. Select > GSP o View – automatic execution of the following functions o GSP Layout > Default/2-3 Segments/Maestro UI/Cardio UI/Tim Planning UI/Spine Planning UI o Defines layout of reference images (2, 3 segments, etc.) o Image text > Default/On/Off o Show (On) or hides (Off) images text o Reference lines > Default/On/Off o Displays reference lines and shows positions of other reference images and facilitates spatial orientation o Protocol table position > Default/On/Off o Displays table position of isocenter protocol as blue line o Coil positions > Default/On/Off o Displays coils and coil elements connected for the examination o Adjust volume > Default/On/Off o Displays adjustment volume in reference images o View-Graphics o Line Mode/Box Mode > Default/Line mode/Box mode o Line Mode – Displays center line of intersection o Box Mode – Displays actual intersection surface and anatomical region covered by slide SIEMENS www.usa.siemens.com/healthcare Page 2 of 3 o Package mode > Default/On/Off o Hides individual intersections of a slice group o Projection > Default/On/Off o Switches on the projection display of slices/slabs o Shadow > Default/On/Off o Displays shadow lines to improve visibility of graphic objects in light images o Image o Windowing on series > Default/On/Off o Activates windowing on complete series o Zoom/pan on series > Default/On/Off o Transfers changes in one image to all images of the respective series o Zoom/pan > Default/Zoom/pan/Zoom/pan + modify graphics o Activates Zoom/Pan mode o Tools o Extent mode > Default/Distance/Number of Slices/Slice Oversampling/Thickness o Activates mode to change extent of slices/slabs (e.g., distance, thickness, slice oversampling) o Extent phase oversampling > Default/On/Off o Activates mode for changing phase oversampling o Coupled graphics > Default/On/Off o Groups graphic objects in order to rotate or move them together o Scroll o Display order >Default/Anatomical/Triggered/as Numbered/as Loaded/Chronological/User-defined o Defines a scheme for sorting images within a series o User-defined – three selection lists are displayed for defining sort criteria SIEMENS Dot Add-In Configurator x t1_f12d_sag EE Copy Paste View Image Guidance GSP layout Default Vindowing on series Default Parameters Image text Default Zoom/pan on series Default Queue Reference lines Default Zoom/pan Default GSP Protocol table position Default Tools Series Label Coil positions Default Extent mode Default Auto load Adjust volume Default Extent phase oversampling Default Auto Coverage View-Graphics Coupled graphics Default Auto Timing Line / box mode Default Scroll Package mode Default Display order Default Projection Default Shadow Default Show Preview Close SIEMENS www.usa.siemens.com/healthcare Page 3 of 3
- MAGNETOM Avanto
- MAGNETOM Aera
- MAGNETOM Skyra
- MAGNETOM Verio