Dimension® EXL™ with LM Integrated Chemistry System Maintenance Online Training
Perform weekly, monthly, cycle-based and periodic maintenance. This clinical laboratory training qualifies for continuing education units (CEU).
Identify periodic maintenance tasks Identify cycle-based maintenance tasks Identify monthly maintenance tasks Identify weekly maintenance tasks Recall the daily maintenance process Welcome to the Maintenance training course for the Dimension ® EXL™ with LM integrated chemistry system. After completing this course, you will be able to: Select Next to continue. Perform Daily Maintenance at least once every 24 hours. Daily Maintenance includes the following tasks: Clean the sample area and empty cuvette waste Check for other maintenance requirements on the Daily Maintenance Routines screen Run a System Check Update the Maintenance log TIP: Refer to the Daily Setup online training course for details on how to perform each maintenance task. Refer to the RMS online training course for additional details on how to perform RMS Daily Maintenance tasks. Select Next to continue. Weekly Maintenance includes the following tasks: Clean the outside of the HM wash probes Clean the outside of the R2 probe and nut Update the Maintenance log TIP: Refer to the Operator’s Guide for details on how to perform each maintenance task. Weekly Maintenance Learn how to perform each Weekly Maintenance task. Checklist TitleChecklist TypeChecklist ContentWeekly MaintenanceHTMLClick each checkbox to learn about the weekly maintenance tasks. Clean HM Wash ProbesHTML Remove any residue by cleaning the wash station probes weekly with water. With the system in Standby, start this maintenance procedure by going to the HM Pump Prime screen to prevent any automatic priming of the HM wash probes. To display the Pump Prime screen, from the Operating Menu, select F4: System Prep, F7: Pump Prime and F6: HM.Clean R2 Probe and NutHTML Perform this task from the HM Pump Prime Screen. Clean the R2 reagent probe weekly using 0.1N sodium hydroxide to remove any residue. Be sure to scrub the nut at the top of the probe during this cleaning procedure. Update Maintenance LogHTML After completing weekly maintenance, update the Maintenance log. Select the X in the upper-right corner to close the window and continue. Monthly Maintenance includes the following tasks: Replace IMT pump tubing Clean the IMT system Stylette the HM wash probes Replace the HM wash station pump heads Clean the R2 and R3 drains Clean and replace instrument air filters Update the Maintenance log Monthly Maintenance Learn how to perform Monthly Maintenance tasks. Checklist TitleChecklist TypeChecklist ContentInstructionsHTMLClick each checkbox to learn how to perform monthly maintenance tasks. Replace IMT Pump TubingHTML The X tubing which stretches around the IMT pump assembly must be replaced monthly to compensate for wearing and stretching, which can affect sample positioning as the pump pulls air and fluid segments through the IMT system. If the media does not automatically start, select the play arrow to begin.Clean the IMT SystemHTML The IMT system tubing and IMT port need to be cleaned with bleach which destroys sensitive electrode membranes in the QuikLYTE® multisensor. NOTE: for demonstration purposes, the fluid shown in the IMT tubing has been colored red for visibility. In addition to monthly maintenance, the IMT QuikLYTE® mutlisensor is replaced every 5 days or 1000 tests. The system will alert you when this maintenance is required. If the media does not automatically start, select the play arrow to begin. Stylette HM Wash ProbesHTML The HM wash probes are styletted to remove any protein build-up from inside the HM wash probes. Do not remove the probes when performing this procedure. Only stylette the aspirate (bent top) probe on each wash probe. If the media does not automatically start, select the play arrow to begin.Replace HM Pump HeadsHTML HM pump heads are changed monthly to compensate for the wearing of the pump tubings. Only the HM pump heads on the wash station are required to be replaced monthly. The shipping plug (or “T”) must be inserted into the pump head before installing onto the motor to prevent damage to the pump rollers as the pump head is pushed onto the motor shaft. If the media does not automatically start, select the play arrow to begin.Clean R2 and R3 DrainsHTML When cleaning the reagent drains, make sure you have a paper towel around the drain to catch any overflow. If the media does not automatically start, select the play arrow to begin.Replace Air FiltersHTML Replace the air filters to ensure air circulating in the instrument is free of dust and debris. There are 3 filters to replace: Cabinet Cuvette ring filter on back (top photo), Power Supply filter beside power switch (not shown) and Thermal Chamber filter (bottom photo). The air flow arrow on the new filter should point IN toward the instrument. The thermal chamber filter is not cleaned, but replaced every month. It does not have any air flow arrows and cannot be installed backwards.Update Maintenance LogHTML Update the Maintenance Log after completing monthly maintenance tasks. Select the X in the upper-right corner to close the window and continue. Cycle-based Maintenance refers to tasks that need to be performed based on a systems cycle counter and include the following: Clean the sample probe and drain Replace the cuvette diaphragm Replace the cuvette film cartridge Replace sampler and reagent probe tips (R1, R2, and R3) To access the System Counters screen, select F4: System Prep, F6: Sys Counters. The System Counters screen also can be accessed from the Daily Maintenance Routines screen when other maintenance is required. Cycle-Based Maintenance Learn how to perform Cycle-based Maintenance. Checklist TitleChecklist TypeChecklist ContentCycle-Based MaintenanceHTML Select each checkbox to learn how to perform cycle based maintenance tasks. Clean Sample Probe and DrainHTML Clean the photometric sample probe and drain every 20,000 samples to remove protein buildup and prevent residue from entering the system and to maintain the precision and accuracy of the photometric sampling system. In the Pos field on the System Counters screen, type the segment position where you will place the Sample Cup filled with undiluted bleach to clean the sample probe. Follow the screen prompts. The sample probe will move up and away from the drain to allow manual cleaning of the sample probe drain. Use a disposable drain brush and a 10% bleach solution to clean the sample probe drain. Use your laboratory‘s biohazard waste disposal procedures when discarding this brush.Replace Cuvette DiaphragmHTML The cuvette diaphragm assists in the formation of cuvettes and should be replaced every 12,000 cycles. When working in the cuvette manufacturing area, be careful not to touch the U-seal solenoid which can be extremely hot. New cuvette diaphragms are packaged with new cuvette film cartridges. Look in the center of the cartridge and do not accidentally throw these away. If the media does not automatically start, select the play arrow to begin.Replace Film CartridgeHTML Replacing the film cartridge ensures the system has enough film to make cuvettes onboard the system. The cuvette film cartridge makes approximately 12,000 cuvettes. The instrument alerts you when there are 300 or fewer cuvettes remaining on the cartridge by displaying an icon in the Operating Conditions Status area. With the new film cartridge in place and the new film correctly aligned and taped to the old film, hanging down from the instrument, select F1: Load Film. Wait for the counter to reset to 12,000 and movement to stop to complete the procedure.Replace Probe TipsHTML Replacing probe tips ensures the accuracy and precision of the sampling and reagent delivery systems. Replace probe tips that have reached a certain cycle count as indicated on the System Counters screen. If the media does not automatically start, select the play arrow to begin. When complete, select the X in the upper-right corner to close the window and continue. These tasks are performed during Periodic Maintenance: Replace Photometer Source Lamp Clean Cuvette Windows Replace LOCI® Insert Replace LOCI® Vacuum Cup Periodic Maintenance tasks Learn about Periodic Maintenance tasks. Checklist TitleChecklist TypeChecklist ContentPeriodic MaintenanceHTMLSelect each checkbox to learn about Periodic Maintenance Tasks. Replace Photometer Source LampHTML Replace the source lamp in the cuvette thermal chamber when the lamp burns out or as part of troubleshooting. If the media does not automatically start, select the play arrow to begin.Clean Cuvette WindowsHTML Clean or replace cuvette windows as needed or when troubleshooting. There are 2 procedures for cleaning cuvette windows: clean all or clean bad (dirty) windows. In the video, the access mode selected is: Access All. If the media does not automatically start, select the play arrow to begin.Replace LOCI® InsertHTML Perform a controlled power shutdown first. Avoid getting fingerprints on new insert. After replacing insert and seal, press firmly down around the entire surface to ensure seal is inserted fully. After restoring power, follow instructions in the Operator’s Guide to perform LOCI® module statistics.Replace LOCI® Vacuum CupHTML From the LOCI® Diagnostics screen, select F6: Inspect to rotate the LOCI®arm out above the HM wheel. Pull down on the vacuum cup to remove it. After replacing, perform LOCI® arm alignments. If the media does not automatically start, select the play arrow to begin. When complete, select the X in the upper-right corner to close the window and continue. Congratulations. You have completed the Maintenance course for the Dimension® EXL™ with LM integrated chemistry system. Listed below are the key points that have been presented. Take time to review the material before you proceed to the final quiz. Recall the Daily Maintenance tasksDaily maintenance tasks consist of cleaning the sample area, emptying the cuvette waste, checking for Other Maintenance requirements on the Daily Maintenance Routines screen, running a System Check and Updating the Maintenance Log. The tasks must be performed at least once every 24 hours. Identify Weekly Maintenance tasks Clean the HM wash probes: Use water to remove residue from HM wash probes. While performing this task, use the HM Pump Prime screen to prevent automatic priming of the wash probes. Prime HM Wash following cleaning. Clean the R2 reagent probe and nut: Use 0.1N sodium hydroxide to remove any residue from the R2 reagent probe and nut at the top of the probe. Identify Monthly Maintenance tasks Replace IMT pump tubing: Replace the X tubing around the IMT pump to compensate for wearing and stretching, which can affect sample positioning across the QuikLYTE® multisensor. Clean the IMT system: Bleach the IMT system when it’s time to replace the QuikLYTE® multisensor and update the count. Bleach destroys sensitive electrode membranes, so be sure to change the QuikLYTE® multisensor AFTER the bleaching procedure has been completed. Clean and replace instrument air filters: All filters can be cleaned and reused except the thermal chamber filter, which must always be replaced. Stylette the HM wash probes: Remove residue from inside the probes. Replace HM wash station pump heads: HM pump heads are replaced monthly because the pump tubings wear and stretch over time. Clean R2 and R3 drains: The drains for reagent probes 2 and 3 are cleaned monthly to ensure effective rinsing of the probes. Identify Cycle-based Maintenance tasks Clean the sample probe and drain: Use undiluted bleach to clean the sample probe, and 10% bleach solution to clean the sample probe drain. Replace the cuvette film cartridge: If 300 or fewer cuvettes remain on the cartridge, an icon appears in the Operating Conditions Status area on the system computer monitor. Replace the cuvette diaphragm: Before removing the cuvette diaphragm, on the System Counters screen, press F2: Diaph Change to de-energize the U-seal solenoid. New cuvette diaphragms are packaged with new cuvette film cartridges. Replace sample and reagent probe tips (R1, R2, R3): Sample and reagent probe tips fit inside sample and reagent probe nuts. Always replace the probe locking nut when you replace a probe tip. After the sample probe tip is replaced, perform a probe alignment. Identify Periodic Maintenance tasks Replace the source lamp: Before replacing the source lamp, be sure to perform a Controlled Power Shutdown. After replacing the source lamp, you must also perform a photometer alignment, mAU offset calibration, run a System Check, run QC, and recalibrate all turbidimetric test methods. Clean cuvette windows: Use the Window Check screen to select cuvette cleaning options, such as cleaning all or only bad cuvette windows, or replacing windows for cleaning at another time. Replace LOCI® vacuum cup: From the LOCI® Diagnostics screen, press F6: Inspect to rotate the LOCI® arm out above the HM wheel. Pull down on the vacuum cup to remove it. After replacing, perform LOCI® arm alignments. Replace LOCI® insert and seal: Perform a controlled power shutdown first. Avoid getting fingerprints on new insert. After replacing insert and seal, press firmly down around the entire surface to ensure seal is inserted fully. Select Next to continue.