Replacing Flex Insert R2 1500/3000T Video

This video provides instructions on how to replace R2 flex inserts on the Dimension Vista® System.

begin cleaning the reagent arm to flex, insert, release the center interlock an lift the center lid. For easier access to the insert, remove the cuvet wash shield. Lift the reagent arm to flex answered out of the instrument to remove it for cleaning. Hold the insert under running water to clean it. And dry it thoroughly with a paper towel or gauze. Reinsert the reagent arm to flex insert, being careful to position in place, tighten the thumbscrew to position properly. Next, carefully reinstall the Cuvette wash shield. Now that you have finished cleaning the reagent arm to flex insert, close the center lid to access the software.

Dimension [email protected] 150013000T System Clean Flex Insert R2

  • monthly
  • maintenance