Resolving Jammed Cuvette at Cuvette Carousel Video
This video provides instructions on how to resolve a jammed cuvette at the Cuvette Carousel on the Dimension Vista® System.
Release the center cover interlock and raised the lid. Lift the cuvet cartridge and look for any gem DCU vets at the bottom of the cartridge. Remove any jammed or broken cubits from the carousel. Confirm there are no straight Q vets on the Cuvet Bridge. Reinstall the cuvet cartridge and ensure it is firmly placed into position. After resolving the jam, closed the instrument lid.
Resolve Jammed 2 Jammed Cuvette at Cuvette Carousel el 707-G 22 707-6
- maintenance