Emptying the Solid Waste on the Atellica® IM Analyzer Video Version 1.25 and Below
This video provides instructions on how to perform the task of emptying the solid waste on the Atellica® IM Analyzer.
This video will show you the steps for how to empty the solid waste on the Atellica Immunoassay Analyzer. The solid waste includes used kovetz and sample tips. The inventory icon on the command bar will alert the user when the waste needs to be emptied. After selecting the inventory icon on the Command bar, select the Supplies overview tab. And then select the immuno assay module graphic. Locate the status of the cuvet and tip waste. A yellow triangle indicates cuvet waste or sample tip waste is greater than or equal to 90% of its capacity or the way status is OK and the operator has removed the waste bin. A red triangle indicates the waist level is full. To empty the solid waste, open the solid waste door. Hold the solid waste drawer out. The drawer will come out completely. Hold on to the drawer as you are moving it to ensure it does not fall. Remove the solid waste bins and empty the contents of the bins into a laboratory approved container for bio hazardous waste to facilitate solid waste bin removal, grasp the bin with both hands and shake slightly. Return the empty solid waste bins to the solid waste drawer. Slide the solid waste drawer into position by properly aligning the drawer. Push the drawer in an close the solid waste door. To confirm the cuvet waste is empty, select yes on the primary display or module display when prompted.
SIEMENS SIEMENS. SIEMENS . AtellicaTM Solution Version I. AtellicaTM Solution Version I AtellicaTU Solution Version 1. .12.0.2777002 12.0.2777002 SP00107 Healthineers System Systan Calltxation Calltxatbn Cahtxation Cahltxation C aux*on Calitr*obn Cahltxatbn Patient Orders Pünt Or&rs P*ient Orders Seno Inventory I nventory Calltxat•on Reagent Overview Supplies Overview Reagent Needs Reagent Loader Reagent Needs Reagent Overview Reagent Needs Reagent Loader Cal-QC Needs Needs Supplies Overview Reagent Overview Reagent Needs Reagent Lot Comparisor. Reagent Lot Comparison Cal-QC Storage Inventory Cal-QC Needs System Fluids Volume Status OBS Lot Expiration Sequence Number IMAcid 1M Acid 1040 27 d2h 27d2h 9/123/2018 9/23/12018 9/23/2018 00114 0018 27 d2h 27d2h &'25.Q018 1M Base 11 0011 1040 1M Cleaner 6 d 22 h 9/23/2018 8/25/2018 00184 0018 25 d 23 h 27 d 23 h 6 d 22 h 8/25/12018 9/25/12018 2223 1M wash 0K 27 d 23 h 25 d 23 h 6 d 22 h 27 d2h 4/25/2018 8/25/12018 1040 Replace System Fluids Supplies Waster Waste Wastem Status 4 Storage Locations Available 4 Storage Locations Avadlable TIPS Cuvette and Tip Waste Cuvette and TIP Waste Greater than 90 percent full 1 TIP Trays In Storage TIP Trays In Storage Cuvettes Tip Tray Waste Tray Waste Open TV Tray Waste Drawer Open Tip Tray Waste Drawer Open Tray Waste Drawer Open Tv Tray Waste Drawer Open Ti Tray Waste Drawer Liquid Waste 0K October 18, 2017 30801 Wednesday, 18, 2017 30801 w«mesday, October 18, 2017 30801 18, 2017 30801 PEP LabManager
- attelica
- attellica
- atelica