epoc® Blood Analysis System with NXS Host Identify Out-Of-Range Results On The Test Card Job Aid
This job aid will show you the steps on how to troubleshoot out-of-range results on the epoc Test Card.
epoc® with NXS Host Blood Analysis System SIEMENS Healthineers Identify Out-Of-Range Results On The Test Card Job Aid Out-of-range results on the Test Card for Liquid QC and Calibration epoc BLOOD ANALYSIS From the epoc Host, disconnect from the Reader, then Run Blood Test View Blood Test ? reconnect. If the wireless connection is successful and Electronic QC passes, verify the following and then repeat the O About test: 1. Use of correct control or Calibration Verification Value Assignment Datasheet O O 2. Use by Date of Controls has not been exceeded 3. The Controls have been handled correctly: see Instructions for Use 4. Test Cards and Controls have been stored correctly ➢ If the repeat results are in range, the cards are acceptable for use. If the results are still out of range despite meeting the above criteria, repeat the test using a new box of control solutions and/or Test Cards. Contact the Siemens Healthineers Remote Services Center if Controls or Test Cards continue to fall outside of specified ranges Template Effective Date: 27 May 2019| HOOD05162003010284 1 Effective Date: 07-JUN-2021| HOOD05162003194856 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2021 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Template Effective Date: 27 May 2019| HOOD05162003010284 2 Effective Date: 07-JUN-2021| HOOD05162003194856 Unrestricted © Siemens Healthineers, 2021
- epoc
- range
- test
- card