epoc® Blood Analysis System with NXS Host Troubleshooting Failed Electronic QC Job Aid
This job aid will show you the steps on how to troubleshoot failed electronic QC on the epoc Reader.
epoc® with NXS Host Blood Analysis System SIEMENS Healthineers Troubleshooting Failed Electronic QC Job Aid Failed Reader Electronic QC ➢ If a Reader fails electronic QC, first confirm the failure epoc BLOOD ANALYSIS ➢ Close the Reader Screen, turn the Reader OFF and ON, and Run Blood Test then try to connect to the Reader again View Blood Test ? ➢ If Reader connects successfully, and therefore passes O About electronic QC, it is acceptable for use. If the problem is not resolved, contact the Siemens Healthineers Remote Services Center O O Template Effective Date: 27 May 2019| HOOD05162003010284 1 Effective Date: 07-JUN-2021| HOOD05162003194856 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2021 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Template Effective Date: 27 May 2019| HOOD05162003010284 2 Effective Date: 07-JUN-2021| HOOD05162003194856 Unrestricted © Siemens Healthineers, 2021
- epoc
- troubleshoot
- electronic
- QC