Guide & Go Job Aid (VA20)
Guide&Go Job Aid SOMATOM go. VA20 This job aid provides hints, tips and tricks for successful Guide&Go set up. Step 1: Adjust brightness setting on tablet to maximum by swiping left to right on the screen. 97% 7 minutes until fully charged Power mode (plugged in): Better performance Best battery life Best performance G0 Battery settings 3. 5 OOK ng. Step 2: If using voice control, enable under settings wheel>assistance systems>voice control. If not using, make sure it is deactivated. Assistance Systems Voice Control Enable voice control commands About Scan&GO X -1212.0 150.0 Scanning not possible -+ Go to Console. O SIEMENS Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 1 of 9 Healthineers Guide&Go Job Aid | SOMATOM go. VA20 Effective Date: 8/22/2019 | HOOD05162003030412 Note: Additional voice control setting should be preconfigured under the tablet settings. Bluetooth devices are supported but have to be configured separately (see Instructions for Use chapter 9.2 CT- guided intervention for full details). Step 3: Register patient and select desired protocol. From the settings wheel, ensure intervention configuration is set. Patient registered × Adult Folder Category Adult Intervention Trauma Thorax Intervention Intervention ~ Abdomen Intervention Dynamic -Indications: biopsy, abscess in the abdomen, w/wo CM -Scan modes: RoutineSpiralAdultAbdomen InterventionSequenceAdultAbdomen2d Topogram Child GranAGO Approved Spine Pain Therapy Settings wheel Back GO -750.0 125.0 0º General Intervention User Position Position of Physician Left or Right Side of the Table O Image Information Gantry Laser Display Laser Crosshair for Measurements Image Text Display White Text in Image Comers SIEMENS Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 2 of 9 Healthineers Guide&Go Job Aid | SOMATOM go. VA20 Effective Date: 8/22/2019 | HOOD05162003030412 Step 4: Perform topogram and planning spiral scan. Abdomen Interventio ... Recon Ranges Topogram --- Sequence Abdomen IP Spedi Abdomen Abdomen Reconstruction Close Contem patient position Carce 14.85 1.25 Step 5: On the tablet, identify the target position by scrolling through the images with the slider. The selected image will be the target position. Patient registered X C Last/Next 3 44 5.15 4 94 165 CTDIvol (32 cmn) [m3Y] CLP [mGY cm Accumulated CTDIvol (32 cm) [mG'] DLP [mGY'om] Intervention, Guide&GO 123455 SOMMATOM go. Top "10/5/1983. F. 34Y 415/2018 HFS 11:18:46.85 AM 2 IMA 29 SPI 200 SP .1,327.7 A Target Position set to -1284.0 mm KV EM. MA5 10¢ Qualiy rel mas 411 110.5 GT DO SL 6.0/64-0,6/00.8 365 -210 Planning 68 Activate scan or finalize intervention Finalize GO -1506.5 125.0 Multiple scans at the same position SIEMENS Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Healthineers Page 3 of 9 Guide&Go Job Aid | SOMATOM go. VA20 Effective Date: 8/22/2019 | HOOD05162003030412 Step 6: Press the ruler icon to make a measurement. Touch screen to set skin entrance and also target locations. Ocm 52 mm F 1120 Step 7: Select Go icon and move button on control box or Remote Scan Control (RSC) to align image target position with gantry laser lights and the patient. Mark the patient and move table out to prep patient for the procedure. Step 8: Place sterile cover on tablet and on RSC (depending on workflow). SIEMENS Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 4 of 9 Healthineers Guide&Go Job Aid | SOMATOM go. VA20 Effective Date: 8/22/2019 | HOOD05162003030412 Step 9: When ready to perform first i-sequence scan, press move button to align patient back into gantry at target position. Press start button (RSC, control box, foot pedal) to initiate scan. Center image is displayed in larger segment and remains the target position. If a different thumbnail is selected (feet, head), it becomes the new target position. Before next scan, press move button and table will move to that new location and i-sequence scan will be centered there when start button is pressed. Patient registered C Last/Next 3.44 "CTDivol (32 cn) [GY] 5.15 DLP [inGY an] Accumulated 8.38 170 CTDivol [32 cm) [mGY] DLP [mGY'com] GOWATCHES TEP 1144 2972: F 450 EraDio Foot Contor Head Continue scanning or finalize intervention Finalizo GO -1262.5 125.0 Press START - Food = 0 O SIEMENS Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 5 of 9 Healthineers Guide&Go Job Aid | SOMATOM go. VA20 Effective Date: 8/22/2019 | HOOD05162003030412 Step 10: Continue workflow and when the procedure is finished, finalize i-sequence mode and end the exam OR perform the post-procedure scan. Continue scanning or finalize intervention Finalizo GO Hints: Gantry can be tilted during exam by graphically tilting the scan range box on the topogram image. If the user requires additional space in the bore for adjusting needles, they can lower the table height using the down arrow on the RSC. The table will adjust and maintain the same table position. Parameters can be changed by selecting the corresponding icon. Window changes are applied immediately. Scan parameter changes will apply to the next i-sequence scan after Go button is pressed to confirm changes. LaxiNal 3.44 5.15 4.04 165 3.44 Abdomon Intorvontion (Adull) Houd First - Supine Abdomen Intervention [Adull) G Houd Fuel - Supino Full 0.33 50 70 Full10:33 50 120 Br 39 HIC 44 100 Hir 48 110 58 120 60 130 64 140 SIEMENS Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 6 of 9 Healthineers Guide&Go Job Aid | SOMATOM go. VA20 Effective Date: 8/22/2019 | HOOD05162003030412 Latest scan is always displayed and selecting the up and down arrows will access previous scans. 5 Feet Center Head Keep in mind that the scan start button will be flashing green regardless of where the patient is because the i-sequence mode stays loaded and can scan at any table position. To move the patient into the position that is shown within the large segment (target position), hold the “move” button. NOTE: The move button will be lit up and static (not blinking like in other workflows). Once you reach the position of the large segment, the move button will become inactive and dimmed on the RSC. If during the procedure you need to go back to the planning spiral series, select double arrow button. Here you can view original measurements, make new measurements, or scroll through planning series and set a new target position. IF Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 SIEMENS Page 7 of 9 Healthineers Guide&Go Job Aid | SOMATOM go. VA20 Effective Date: 8/22/2019 | HOOD05162003030412 Ability to magnify the area of interest during a measurement by touching and holding on the screen. Modify measurements by selecting the line and moving the blue cross marks. To delete a measurement hold down on the line until trash can appears and drag into the icon. 10/5/1983, F, 34Y 415/2018 11:18:45.92 AM R SIEMENS Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2019 Page 8 of 9 Healthineers Guide&Go Job Aid | SOMATOM go. VA20 Effective Date: 8/22/2019 | HOOD05162003030412 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (herein- after collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2019 Local Contact Information Siemens Medical Solutions USA, Inc. 40 Liberty Blvd. Malvern, PA 19355-9998 USA Telephone: +1-888-826-9702 siemens.com/healthcare siemens.com/healthineers ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 9 of 9
- Guide & Go Job Aid (VA20)