How to add and configure new users - MAMMOMAT Revelation
Use the following step by step to add and configure new users on the MAMMOMAT Revelation.
How to Add and Configure New Users – MAMMOMAT Revelation MAMMOMAT Revelation Use the following steps to add and configure new users on the MAMMOMAT Revelation. The new user will be created in the Security Configuration and then the users’ preferences will be config- ured in Procedure Handling. This will need to be done for each new user and each bilateral RIS procedure. Unilateral RIS procedures are created for ‘all users.’ 1. Creating New User: a. From the top drop-down menu, click Options, Configuration, Security, and then Con- figure b. Select User Management. c. Right click on Users, New, then User synge Name User Management Zi Users Icers xi!Internal Users Jew User >i! Internal Users >i- DICOM Nodes si DICOM Nodes Xi Groups Help *# Groups > Roles *8 Roles @ It Security Managemen d. User tab: aygo Sucurity Configuration i. Name- tech’s username. This is not case sensitive ii. Full Name- tech’s first and last User * Users name - iii. Description- Mammographer 4 % Eecurity Mmagene: Namo iv. Password- cannot be any part Full nam of the tech’s username or full name. v. Password Never Expires- Neerpassword checked Confirm password vi. Click Apply e. Member Of tab: Account is disabled i. Add Administration ii. Add Security Admin User cannel change password iii. Click Apply f. Owner Of tab: Cuncol HOID i. Add Role Doctor ii. Add Role MTA iii. Click Apply g. Close security configuration and end session. SIEMENS Healthineers 2. Configuring User Preferences: Options Help JE WH AWS - Configuration Panel Switch to Operating System Login. File a. Log in as new user b. From the top drop-down Maintenance .. End Session menu, click Options, Config- Auto Enable First PV WH AWS MoodLight Patient landing C. Transfer C Browser forcate Screen Saver uration, then Procedure Handler Configuration ... Handling Service ME Security Add Printer Transfer Audit Trail Filming Study Filming Layout Layour Layout Editor c. Select the RIS procedure with a 1 in Procedure Handling Configuration Procedures EXPOSURE Programs front, ex. 1Screening Bilateral Tomo + Projection names 2D PRIME. List of Procedures Procedure details Bilateral Mammogram+Spot Name Bilateral Mammogram - PRIME d. Click COPY (button at bottom of R Mammogram L Mammogram Examinant on purpose Diagnosis screen, down below List of Proce- R Mammogram+Mag L Mammogram+Mag Store procedure as R Mammogram+Spot dures box) L Mammogram+Spot Show Procedure Code Magnification Spot Implant Bilateral Implant+Implant Displaced Bik F Auto Decomp F Opcomp Constancy Test Calibration OC-raw QC-processed Default Display Layout 1:1 Biopsy Calibration BiopsyLeft Autofilming Autoexpose Skip InExam Transfer BiopsyRight BiopsyLeft Tomo BiopsyRight Tomo IP SO Flavor BiopsyLeft Tomo Plus BiopsyRight Tomo Plus Display Markers Except mag/spot Bilateral Tomo Bilateral Tomo + 2D Bilateral Tomo + 2D PRMME QC-raw Tomo QC-processed Tomo Projection View Set Bilateral Mammogram - PRIMI R-CC PRIME L-CC PRIME R-MLO PRIME L-MLO PRIME AB_Blateral Mammogram AB_Bilateral Torno New Copy Dologo Sot derau Modify PVS OK Cancel Appy e. Delete all the info that is not part of the RIS exam Copy of 1.Then change Examination purpose to either Screening or Diagnosis. Change Store procedure as to current user only. Procedure details Name Screening Bilateral Tomo + 2D PRIME Examination purpose Diagnosis Store procedure as current user only Show Procedure Code 3. Check the box for ‘Skip InExam Transfer’. Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2022 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 01/04/2022 | HOOD05162003247432 Healthineers Page 2 of 6 Auto Decomp Opcomp Unprocessed Exposure Program set QC_set2 Default Display Layout 11:1 Autofilming Autoexpose Skip InExam Transfer a. Navigate to Modify PVS. b. Click on the Modify PV Sequences tab, then change the view se- quence by clicking and dragging the projection views listed under Selected PV se- quence, in the desired order. c. Click Apply, then OK. Procedure Handling Configuration List of Procedures Procedure details Bilateral Mammogram.Spot Name Bilateral Mammogram . PRIME R Mammogram L Mammogram Examination purpose Diagnosis R Mammogram+Mag L Mammogram+Mag Store procedure as R Mammogram+Spot L Mammogram+Spot Show Procedure Code Magnification Spot Implant Bilateral Auto Decomp F Opcomp Implant+Implant Displaced Bill Constancy Test Calibration QC-raw QC-processed Default Display Layout 1:1 Biopsy Calibration BiopsyLeft Autofilming Autoexpose Skip InExam Transfer BiopsyRight BiopsyLeft Tomo P Set BiopsyRight Tomo PVS Dialog BiopsyLeft Tomo Plus BiopsyRight Tomo Plus Display Markers Except mag/spot Bilateral Tomo Modify PVa Modify PV seguprices Bilateral Tomo + 2D Bilateral Tomo + 2D PRIME QC-raw Tomo Projection View set QC-processed Tomo Current Projection View sel Bilateral Mammogram - PRIMI R-CC PRIME L-CC PRIMER-MLO PRIME L-MLO PRIME AB_Blateral Mammogram AB_ Bilateral Tomo R-CC TOMO L-CC TOMO L-MLO TOMO H-MLO TOMO New Copy Delose Set defaut Modify PVS OK Cancel Apply Acquisition mode Tomo Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2022 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 01/04/2022 | HOOD05162003247432 Healthineers Page 3 of 6 d. Click ‘Apply’ after each proceure that is modified. Complete for ALL bilateral RIS pro- cedures. L-FR R-FB L-SIO R-BIO OK Cancel Apply 4. After Procedures Are Configured: a. Select the screening mammogram pro- cedure and click ‘Set Default.’ NOTE: This procedure will be used if a procedure is registered that has not been configured and/or mapped with the RIS. b. Click Apply. Later in Diped c. Uncheck all clinical factory exams as well as the following: i. Ex. Bilateral Tomo, Bilateral Ma- BkpirSon Trewit mogram, Magnification, Spot, Implant Bilateral, etc. ii. Biopsy exams, Constancy test, IQ d. Move the QC exams to top of List of Procedures by dragging them up or us- ing the up/down arrows: i. QC-raw ARTIFACT DETECTION ii. QC-raw SNR/CNR Conoci iii. QC-raw PHANTOM iv. QC-processed TOMO PHANTOM v. QC-raw COMPRESSION FORCE TEST e. Arrange the exams in order under the QC exams by dragging them up or using the up/down arrows. f. Click ‘Apply’ after all changes Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2022 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 01/04/2022 | HOOD05162003247432 Healthineers Page 4 of 6 5. Change the angle, if different from 45° deg, Projection hames in the Projection names: a. Select MLO from list of preset angle List of Preset Angle configurations configurations Vowtypo Onginal Moditod b. Change preset angle to desired angle ML 900 c. Click ‘Apply’ 450 LM LMO 45.0 45.0 FB 180.0 1800 SPEC 00 XCC XCOL 0.0 00 0.0 00 XCOM 0.0 00 00 45.0 45.0 Reso Presst Argle 45 C : L-MLO: +45 C. R-MLO: -450) Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2022 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 01/04/2022 | HOOD05162003247432 Healthineers Page 5 of 6 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Local Contact Information Legal Manufacturer Siemens Healthcare GmbH Siemens Medical Solutions USA, Inc. Siemensstrasse 1 Henkestr. 127 91052 Erlangen, Germany 40 Liberty Blvd. Forchheim, Germany 91301 Phone: +49 9131 84-0 Malvern, PA, USA 19355-9998 Phone: +49 9131 84-4746 siemens-healthineers.com Telephone: +1-888-826-9702 siemens.com/healthcare siemens.com/healthcare Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2022 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 01/04/2022 | HOOD05162003247432 Healthineers Page 6 of 6
- MAMMOMAT Revelation
- women's health
- configure new users