How to Restore Siemens Default Scanner Protocols

How to Restore Siemens Default Scanner Protocols step by step

How to Restore Siemens Default Scanner Protocols Computed Tomography 1. In this example, clinical protocol parameters were saved in place InnerEarUHR of the Siemens default values for the InnerEar protocol. Note the InnerEar InnerEarUHRSeq Sinus bullet and upper/lower case format of the default protocol. 2. To restore, load the Scan Protocol Assistant (SPA) from Options > Configuration. Click the radio button next to Restore scan Restore scan protocols to Seemens default protocols to Siemens default. Click Step 2 at the top of the screen to advance to the next step. 3. Place a checkmark next to the default protocols to be restored or Replace customized protocols with same name mark the Select all box. Select Add Siemens protocols as O Add Siemens protocols as duplicates duplicates (to preserve the clinical protocol settings and later O Adut correct the protocol name). These steps may be repeated for the Child Child directory if needed. Select all 4. Click Yes to save the restored scan protocols. Exit the SPA and Confirmation return to the Exam card. Create a test patient to display the Patient Model Dialog. Restored scan protocols. Head (Adult) InnerEar 5. In the Patient Model Dialog, confirm the clinical protocol appears InnerEer Customized as its name followed by the word Customized. This protocol ·HEAD ROUTIVE contains the parameters which were mistakenly saved over the Hoad_XCARE_SAFIRE HeacRouine Siemens default protocol. The name can be edited in the SPA to reflect the routine clinical naming style, such as INNER_EAR. HeadSoo_XCARE 6. In the Patient Model Dialog, confirm the Siemens default protocol now appears without a bullet. This restored protocol will appear InnerEarUHR InnerEar with its original parameters in the same alphabetical order InnerEarUHRSeq Sinus among the other default scanning protocols. To preserve the accuracy of Siemens default protocols, save clinical protocols with unique names in all capital letters. SIEMENS siemens.com/healthineers Healthineers . Job Aid | How to Restore Siemens Default Scanner Protocols Effective Date: 8/16/2016 | Quality HOOD05162002267010 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright @ Siemens Healthcare GmbH ...... ........... Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 2 of 2

  • as
  • flash
  • emotion
  • open
  • spirit