Inline Composing - Eline
This Job Aid reviews the workflow steps to manually compose images with multiple table moves on MR Systems with Eline software.
Inline Composing Eline Job Aid Inline Composing – Eline This Job Aid will review workflow to facilitate planning and evaluating multi-slice measure- ments, group the individual measurements or combine examinations with multiple table moves using Inline Composing. Prerequisites for Inline Composing Distortion Correction – On, automatically set to 2D Distortion Correction (e.g., Inline > Composing Parameter Card) Positioning Mode – ISO or FIX Mode must be selected (e.g., System > Miscellaneous Pa- rameter Card) Routine Contrast Resolution Geometry System Physio inline Sequence Common MIP Composing Distortion Corr. v Mode 2D Unfiltered images Inline Composing v Normalize None Composing Function Spine Composing Group Last Step Series Description Composing Group* 100 Routine Contrast Resolution Geometry System Physio Inline Sequence Coils Miscellaneous Adjustments Adjust Volume TX/RX Coil Combine Mode Adaptive ... Positioning mode ISO Save uncombined Table position F FIX mm ISO Table position memory Matrix Optimization Off MSMA S-C-T Coil Focus Flat Sagittal R>> L Coronal A >> P AutoAlign Spine > Thoracic Transversal H>> F Coil Select Mode On - Auto ... Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 07/29/2020 | HOOD05162003109364 Healthineers Job Aid | Inline Composing - Eline Effective Date: 07/29/2020 | HOOD05162003109364 1. Create a Inline Composing Protocol (e.g., Whole Spine) a. Open Dot Cockpit Icon b. Program Editor > Edit i. Open Localizer (e.g, T1, T2, etc.) c. Select Inline > Composing i. Select Composing Function – Spine (e.g., Spine, Angio, Adaptive, Diffusion) ii. Select Composing Group – Identifies Step being Composed 1. Select Step 1 for C-Spine Localizer iii. Last Step – last Step of a Composed Group (e.g., L-Spine) 1. All Steps for group number defined compose once last step is completed iv. Enter Series Description (e.g., C-Spine, T-Spine, L-Spine) v. Open Next Localizer (T-Spine, L-Spine) and Repeat Steps for each sequence (e.g., T1, T2, etc.) vi. Save Program As – when finished building entire protocol EXPLORER PROGRAM EDITOR Browse Import Export SIEMENS SIEMENS > whole-spine >> clinical libraries » inline_composer Edit _localizer_H_0mm 00:17 i-spine Isnine Il localizer F 220mm 00:17 whole-spine Clinical libraries Ill_localizer_F_430mm 00:17 general lesion inline composer set-n-go |_12_tse_sag_384 › TimCT whole-spine neck soft tissue Il t2 tse_sag_448 04:21 thorax heart breast III_12_tse_sag_448 04:21 abdomen pelvis female pelvis male shoulder I_t1_tse_sag 384 04:53 elbow wrist Il t1 tse sag_384_p2 04:40 hand hip Routine Contrast Resolution Geometry System Physio Inline Sequence knee IlI t1 tse sag_384_p2 Common MIP Composing ankle Distortion ComT. v Mode 2D Unfiltered images Inline Composing Normalize None Composing Function Spine Composing Group 3 Last Step v Series Description L-Spine Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 07/29/2020 | HOOD05162003109364 Healthineers Page 2 of 4 Job Aid | Inline Composing - Eline Effective Date: 07/29/2020 | HOOD05162003109364 2. Inline > Composing Parameter Card a. Inline Composing – combines all images from multiple table moves (steps) for an orien- tation or weighting into a complete image (e.g., C, T, and L spine combined) b. Normalize – Off (Default) – reduces image intensity variation for whole body diffusion measurements (e.g., used for composed whole body diffusion) c. Composing Function – establishes method used to composed images i. Spine – based on Bone Structure, for Composing Spine Images ii. Angio – based on Vessel Structure, for an Overview of Vessels iii. Adaptive – based on elastic matching, to correct b0 induced inhomogeneities in image areas beyond magnetic homogeneity volume. e.g. Abdomen, Long Bones iv. Diffusion – correct phase encoding direction shifts caused by different adjust- ment parameters of steps of a set-n-go protocol (e.g., Axial Whole-Body Diffu- sion) d. Composing Group – Identifies what composing group protocol belongs i. Composes all steps into an overall image (e.g., Whole Spine) e. Last Step – chronologically last measurement step of a Composing Group (e.g., L-Spine) f. Series Description – Name of Composed Step Routine Contrast Resolution Geometry System Physio Inline Sequence Common MIP Composing Distortion Corr. v Mode 2D Unfiltered images Inline Composing Normalize None Composing Function Spine Composing Group 3 Last Step v Series Description L-Spine Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 07/29/2020 | HOOD05162003109364 Healthineers Page 3 of 4 Job Aid | Inline Composing - Eline Effective Date: 07/29/2020 | HOOD05162003109364 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 07/29/2020 | HOOD05162003109364 Healthineers Page 4 of 4
- Inline
- Composing
- Eline