PEPconnect

Inspecting the Washer Probes on the Atellica® CH Analyzer Video Version 1.25 and Below

This video provides instructions on how to perform the daily maintenance task of inspecting the washer probes on the Atellica® CH Analyzer.

The Atellica Chemistry Analyzer has two sets of washer probes that clean the Q vets. The washer probes need to be checked on a daily basis and cleaned if needed. To check the washer probes, the system needs to be set to diagnostic state. This can be done by selecting system in the command bar. Next, select operator diagnostics. Select the Ch module. Next, select enter diagnostics. Finally, select yes. In subsystems, select inspect and clean. Next, select inspecting the washer probes. Select perform. Finally, select yes. Open the back cover and check the washer probes for dirt and buildup. If there is residue buildup, cover the qubits with a paper towel and wipe the washer probes with lint free tissue dampened with special reagent water. Remove the paper towel and close the back cover when the procedure is complete. To complete the procedure, first select continue in the software. Next select close. Select exit diagnostics. Select yes select the maintenance tab in the command bar. So like schedule. Select the module in the drop down menu. Select inspecting washer probes. Finally select the mark is completed button in the software. This completes the overview of inspecting the washer probes.

SIEMENS . Atellica' AtelLicae Atellicae Solution Version 1. sp00107 Version 1. .14.0.2983003 14.0.2983003 14.0.2951003 HeoLtnineers SIEMENS . HeoLtnineers I ry Software Enter Diagnostics Operator Diagnostics Diagnostic Name Diagnostics O i agnostics 1M Failter Faipleted Module: CH Frequency: Al Time: All Performed BY'. All Due Activities: Operator,Automatit Module: CH Frequency: Al Time: All Performed All Due Activities: Operator,Automatit System Module Module Activity sun.30 Mon-Ol Tue.02 Wed.03 Fri.os Sat.06 OiMostic Name Piagnosoic Name Diagnostic Name O Diagnostics Diagnostics Name oescription Description All IMT Caity Clearüng Confirmation Marx as Mar* as Standby Cleaning the Reagent Loader Cleaning the Reagent Cleaning the Reagent Louse' Cleanim the Washer Cleaning the CHagent Cleanin the VVasher Cleanim the Washer Protes Clearung the Reagent Under the Loader the Reagent Louse' the the Loader the Re"ent Loader Reagent Reagent Loader Perform All c-P1-05 Time Inspe:ling Washer Probes Inspe:ting the W'asher Probes Inspe:ting Washer Probes Insß-cting Probes Inspeting the Washer Probes the Washer Prot*s 'hasher Probes washer Probes 'he Washer Probes Probes sco Cleaning the System Fluids Crawer Cleanim the Washer Cleaning the System Fluds Crawer Cleaning the Reagent Louse' Cleaning the System Fluids Cræwer Cleaning the System Fluds Cra•.ver Cleaning the System Fluds Cræwer Cleanim the Washer Protes Cleaning the System Cræwer Cleaning the System Fluds Craw.ver the System Fluids Drawer Cleaning the System Fluids Crawer Cleaning the System Fluds Cræwer the System the System Fluids Confirmation performed By Capabilities CH Dai* Maintenance Cleaning Protps Cleanim the Washer Inspecting Washer Probes Cleaningthe Washer Probes Cleaning Washer Cleanin the VVasher Inspect the Washer Probes Cleaning Washer Cleanin the VVasher Cleaning Washer Prot*s Cleaning W'asher Cleaning Washer Probes Cleaning W'asher Prot*s Inspecting W,'u'asher Probes Cleaning Washer Cleanim the Washer Protes the Washer Probes the Washer the Loader tn the Washer Prot*s Probes the Search Text Clspeaning Washer Probes Inspect the Wash Probes: Insß-cting Prc&s Inspect the Washer Probes: Inspectinge Washer Probes: Cleanim the Washer Protes Inspect the Wash Probes: tn the Washer Prot*s Insß-cting Prc&es Inspecting Probes Inspecting 'Clasher Probes Insß-cting Probes Ins-æcting Prc&s Insg:ectJng Probes Insßecting Probes Inspecting the Washer Probes Inspecting Washer Probes the Washer Prot*s the Washer Scheduled by Operator Washer 1. Refer to the online help for the procedure: Inspecting CH Wash Cleaning and Inspecting the CH Probes Wand Cleaning Ins•æcting the CH and Wxer and the CH and Mixer Impelers Cleaning and Inspecting the CH Probes and Mixer Impellers and the CH and Mixer Irnpelers Cleaning Ins•æcting the CH and Wxer Imßllers and the CH and Mixer Cleaning Insß-ctjng the CH and Wxer and the CH and Mixer Impelers Cleaning Insß-ctjng the CH andt* Cleaning the CH Probes and Cleaning and Ins•æcting the CH and Wixer and the CH Probes and Mixer Imnpelers Cleaning and Ins•æcting the CH and Wxer Imßllers and the CH and Mixer I Cleaning Ins•æcting the CH Probes and Wxer and the CH and Mixer Imnpelers Cleaning the CH Probes and Wxer and CH and Mixer Irnßlers Cleaning the CH and Wxer and CH and Mixer Cleaning and Ins•æcting the CH and Wxer Imßllers and the CH and Mixer Cleaning and the CH Cleaning the CHagent Cleaning Washer Cleaning the CH and Cleaning the CH Wxer Impellers and the CH Probes and Mixer ImpeUers Cleaning Ins•æ-cting the CH Prot* Wash Stations Scheduled Automatically Cleaning Ins.ß-cting the CH Prot* Wash Stations Cleaning and Insgecting the CH Wash Stations Cleaning Ins.ß-cting the CH Probe Wash Stations Cleaning and Insgecting the CH Probe Cleaning and Ins.ß-cting the CH Prot* Wash Stations Cleaning ard Ins*æcting the CH Probe Wash Stations Cleaning and Ins•æcting the CH Prot* Cleaning Ins-R-cting the CH Cleaning ard Ins*cting the CH Probe Wash Stations Cleaning Ins•æcting the CH Cleaning Ins•æcting the CH Prot* Cleaning Ins*cting the CH Probe Wash Stations Cleaning Ins•æcting the CH Prot* Wash Stations and the Probe Wash Stations and Probe Wash Stations Scheduled by Servxe 2. After completing he procedure, select Continue. Connected to FCC Faipleted Failed ,OveIdue Overdue Insß-æct ant Clean Cleaning Insßæct ant Clean Insß-æcting Prc&s Inspect ant Clean Lids Cancel Close I nprocess Continue Yes No Completed Campleted Back Cover Front Cover Not Performed Front Cover Back Cover With Comments Eniter Diagnostic s Exit Enter Diagnostics Exiter Diagnostics Exit Diagnostics Exiter Diagnostic s Enter Diagnostics Exit Diagnostic s Eniter Diagnostics Add Set PEP Wednesday. 2018 Wednesday. 03, 2018 Wednesday. 2018 1:4323 Wednesday. 03, 2018 1:4323 Wednesday. 03, 2018 21052 Wednesday. 03.2018 1:4323 Wednesday, 03, 1 43 23 03, 2018 1:4323 03, 2018 1:14323 03, 2018 2: 10:52 03, 2018 1.14323 Tuesday. July 10.2018

  • attelica
  • attellica
  • atelica