MAMMOMAT Inspiration: How to Activate or Deactivate Display Preview SBS
MAMMOMAT Inspiration: How to Activate or Deactivate Display Preview step by step
MAMMOMAT Inspiration: How to Activate or Deactivate Display Preview Women’s Health/Mammography 1. You must be active in an examination to perform this function. 2. On the examination task card go to "View" on the top tool bar. View Image Tools @ 1:1 b 2.1 c4:1 d9.1 @ 16:1 3. Select or deselect "Display Preview." Display preview Display targeting aid SIEMENS siemens.com/healthineers Healthineers Job Aid | MAMMOMAT Inspiration: How to Activate or Deactivate Display Preview Effective Date: 8/15/2016 | 454 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 2 of 2 © Siemens Healthcare GmbH, 2017
- MAMMOMAT Inspiration PRIME Edition