MAMMOMAT Inspiration VB60 or Higher: How to Add a 2D Procedure
MAMMOMAT Inspiration VB60 or Higher: How to Add a 2D Procedure step by step
MAMMOMAT Inspiration VB60 or Higher: How to Add a New 2D Procedure Mammography/Women's Health 1. From the top dropdown menu, click OPTIONS->CONFIGURATION. Options Help Maintenance ... End Session A Auto Enable First PV Configuration ... Service Security 2. Double click on the "Procedure Handling Configuration" icon. Procedure Handling Configuration 3. Select the appropriate exam and click "Copy" (ex. Bilateral Mammogram-PRIME) List of Procedures QC RAW *MAMMO SCREENING DIGITAL IMPLANTS Bilateral Mammogram Bilateral Mammogram+Mag Bilateral Mammogram+Spot R Mammogram L Mammogram R Mammogram+Mag L Mammogram+Mag R Mammogram+Spot L Mammogram+Spot Magnification Spot Implant Bilateral Implant+Implant Displaced Bilater Constancy Test Calibration QC-raw QC-processed IQ KK KK K Biopsy Calibration BiopsyLeft BiopsvRight Bilateral Mammogram - PRIME Cupy VI DIlateral IVIanImugram - r New Copy Delete Set default SIEMENS siemens.com/healthineers Healthineers Job Aid | MAMMOMAT Inspiration VB60 or Higher: How to Add a New 2D Procedure Effective Date: 1/19/2018 | HOOD05162002781236 4. Fill in "Procedure Details" -name (from your RIS, Ex. SCREENING MAMMOGRAM BILATERAL); -purpose (screening or diagnostic); store as either "current user" or "all users. Procedure details Name Copy of Bilateral Mammogram - PRIME Examination purpose Diagnosis Store procedure as Available for all users 5. Continue to fill in "Procedure Details" with Auto Decomp: Check box, OpComp: Check box, Default Display Layout- 1:1, Auto filming & Auto expose: leave unchecked, Display markers: click on dropdown and select appropriate choice Auto Decomp & Opcomp Default Display Layout 1:1 Autofilming Autoexpose IP Set Flavor Display Markers Except mag/spot 6. Click "Modify PVS" Modify PVS 7. Select the appropriate "Acquisition mode" for the specific views you are adding-Example: PRIME select PRIME, Implant in place, select IMPLANT, Implant displaced, select Standard 2D or PRIME. Select the appropriate PV's and any modifiers needed. Click "OK" Modified Views Please select the view position and modifier(s). Acquisition mode Standard 2D Standard 2D Spot Implant Axillary Tail Cleavage Prime Magnification Rolled Inferior Rolled Lateral Specimen TOTTIO Magnification 2D+Tomo Prime+Tomo L-ML L-SPEC Tomo Implant 2D+Tomo Implant New projection view. L-MLO .- XCC R-XCC R-MLO L-LM L-XCCL R-XCCL R-LM L-LMO L-XCCM R-XCCM R-LM OK Cancel L-CC R-CC L-FB R-FB Siemens Healthineers Services Page 2 of 4 Job Aid | MAMMOMAT Inspiration VB60 or Higher: How to Add a New 2D Procedure Effective Date: 1/19/2018 | HOOD05162002781236 8. Click "Apply" again to populate procedure, then click "OK" Projection View set R-CC PRIME New Copy Delete Set default OK Cancel Apply Siemens Healthineers Services Page 3 of 4 Job Aid | MAMMOMAT Inspiration VB60 or Higher: How to Add a New 2D Procedure Effective Date: 1/19/2018 | HOOD05162002781236 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright @ Siemens Healthcare GmbH ...... Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 4 of 4