MAMMOMAT Revelation Stereotactic Biopsy: How to Release a Patient if Movement is Blocked
How to Release a Patient if Movement is Blocked During a Stereotactic Biopsy MAMMOMAT Revelation In the event of a power failure or if movement is blocked during a stereotactic biopsy examination, release the patient as follows: Removing the needle: 1. If the biopsy device is large and at risk of colliding with the collimator during the removal, then: Loosen the fastening screw (1) of the needle holder. (1) 2. Remove the biopsy device (2) plus the needle holder and needle by slightly tipping the biopsy device (e.g., Mammotome) in the direction of the patient. (2) SIEMENS siemens.com/healthineers Healthineers Job Aid | How to Release a Patient if Movement is Blocked during a Stereotactic Biopsy Effective Date: 03/18/2019 | Quality HOOD05162002998341 Press the compression emergency unlocking device: 1. If the distance between the compression plate and the positioned biopsy unit is greater than 3 cm, then: > 3 cm 2. Firmly press the compression emergency unlocking device (2). Compression is mechanically unlocked, compression plate can be pushed up. – The patient may be removed from the biopsy unit. – (2) 3. If the distance between the compression plate and the positioned biopsy unit is less than 3 cm, then you must first move the needle positioning part up: < 3 cm 4. Using a stylus, open the small cover (4) at the top of the needle positioning part (3). (3) (4) Siemens Healthineers Services Page 2 of 4 © Siemens Healthcare GmbH, 2017 Job Aid | How to Release a Patient if Movement is Blocked during a Stereotactic Biopsy Effective Date: 03/18/2019 | Quality HOOD05162002998341 5. Using the calibration phantom stylus (6) or a ballpoint pen (7), for example, turn the wheel (5) clockwise. The needle positioning part will move upward. – (6) (7) (5) 6. Now firmly press the compression emergency unlocking device (2). Compression is mechanically unlocked, compression plate can be pushed up. – The patient may be removed from the biopsy unit. – (2) O Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthineers Services Page 3 of 4 © Siemens Healthcare GmbH, 2017 Job Aid | How to Release a Patient if Movement is Blocked during a Stereotactic Biopsy Effective Date: 03/18/2019 | Quality HOOD05162002998341 Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 4 of 4 © Siemens Healthcare GmbH, 2017
- MAMMOMAT Revelation Stereotactic Biopsy: How to Release a Patient if Movement is Blocked