PEPconnect

MAMMOMAT Inspiration VB60 or Higher How to Perform a New Biopsy Unit Referencing - Error 7

MAMMOMAT Inspiration VB60 or Higher How to Perform a New Biopsy Unit Referencing - Error 7

How to Perform a New Biopsy Unit Referencing- Error 7 MAMMOMAT Inspiration VB60 or higher If the biopsy unit is mounted on the MAMMOMAT Inspiration, the needle holder automatically moves through the reference positions. The referencing movement of the needle holder must not be interrupted. If a malfunction occurs during the referencing movement, the following window appears: WH AWS X X The Biopsy Unit encountered a hardware problem during self test. Error: 7. A new Biopsy Unit referencing is required. Please remove needle and paddle and push OK to open the user interface for Biopsy Unit referencing. After successful referencing the system should be restarted. OK 1. Click on OK. OK 2. The following window is displayed: @Biopsy Referencing Tool - Microsoft Internet Explorer provided by Siemens Medical So ... x Elle Edit View Tools Help Back > > Search For Address @] C:\WS\utils\loader.htm Go |Links >>> To help protect your security, Internet Explorer has restricted this file from showing active content that could X access your computer. Click here for optio Allow Blocked Content ... An error has occured while trying to access Biopsy referencing p: What's the Risk? Information Bar Help 3. Select the upper line "To help protect..." To help protect your security, Internet Explo access your computer Click here for option SIEMENS siemens.com/healthineers Healthineers . Job Aid | How to Add a New 2D Procedure Effective Date: 04/27/2018 | Quality HOOD05162002869507 4. Open the popup menu (right mouse button) and select Allow Blocked Content. Allow Blocked Content ... What's the Risk? Information Bar Help 5. The following window is displayed: Security Warning X Allowing active content such as script and ActiveX controls can be useful, but active content might also harm your computer. Are you sure you want to let this file run active content? Yes No 6. Click Yes .................................... Yes .............................................................................. 7. The following window is displayed: Zigma Service Modul - Microsoft Internet Explorer provided by Siemens AG Medical Solutions Zigma Webserver Page sent at 10:37:35 07.08.2009 System Version VC00D State E STATE BLOCKED Biopsy Service Please check following: Biopsy is plugged - System is in state where movements (compression unit, rotation, ... ) are possible - No paddle mounted No obstacle on the biopsy table Start Referecing 8. Follow the instructions for manual assignment of the needle holder reference position. 9. Click the Start Referecing icon. Start Referecing The referencing of the needle holder is started. The needle holder automatically moves through several positions. The needle holder should stop moving in the upper left-hand position. Siemens Healthineers Services Page 2 of 4 © Siemens Healthcare GmbH, 2017 Job Aid | How to Add a New 2D Procedure Effective Date: 04/27/2018 | Quality HOOD05162002869507 10. The following window is displayed: http://zigmamu/biopsy/aws_wait_referencing.htm - Microsoft Internet Explorer provided by Siem ... Zigma Webserver Page sent at 10:39:01 07.08.2009 System Version VCOOD State E_STATE_BLOCKED Back to Home Service Mode Ok: confirm the Reference Position. The needle guide should be in the top left inside position Continue: Reference Position is not ok Retry: Move to the reference position again OK Continue Retry 11. Click Continue if the needle holder is not located in the upper left position OR Continue 12. Click OK if the needle holder is located in the upper left position. OK 13. The referencing of the needle holder was successful. The following window is displayed: @ Zigma Service Modul - Microsoft Internet Explorer provided by Siemens AG Medical Solutions Zigma Webserver Page sent at 10:42:53 07.08.2009 System Version VC00D State E STATE BLOCKED Back to Home Referencing done! 14. Click the right button to close the window and the application. x 15. The biopsy unit is ready for operation. Siemens Healthineers Services Page 3 of 4 © Siemens Healthcare GmbH, 2017 Job Aid | How to Add a New 2D Procedure Effective Date: 04/27/2018 | Quality HOOD05162002869507 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 4 of 4 © Siemens Healthcare GmbH, 2017

  • MAMMOMAT Inspiration VB60 or Higher How to Perform a New Biopsy Unit Referencing - Error 7