Manually Removing IM Reagent Packs from the Gripper Video Version 1.25 and Below

This video will explain how to manually remove IM reagent packs from the gripper on the Atellica IM Analyzer.

To remove an I am reagent pack from the gripper on the atellica I am analyzer. The system must be placed into stop state to stop the system. First select system on the command bar. Next select status. Then select the Atellica Immunoassay analyzer icon, select the pause button. This will allow the system to complete any current tasks and wait for processing tests to complete. The instrument will transition from pause to stop. Next, select the button to unlock the front or back cover according to the gripper location. Open the cover. Once the system is paused, you must open the cover within 30 seconds of unlocking, otherwise, the cover will relock to release the I am reagent pack from the gripper. First grasp the bottom of the pack firmly, then move the pack lengthwise toward the center of the system. Rotate the back of the pack downward and out of the gripper. Do not pull a pack out sideways from the gripper. Pulling a pack out sideways results in contamination of the area or disruption of the gripper orings to complete the procedure. Close the cover. Lastly, select the resume button in the software. If the pack is undamaged, reload the pack onto the analyzer.

SIEMENS Solution Atellica HeoLtnineers Please note that the learning material is for training purposes only! Operator Diagnostics For the proper use Of the software, please always use the Operator Manual or Instructions for Use_ (hereinafter collectively "Operator Manual") For the proper use ot the software, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. System c.P1-06 Pl.060 PI-060 Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time Of the training The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Manually Remouing 1M SIEMENS • HeoLtnineers Capabilities Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Atellica Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your Local SIEMENS Siemens Healthineers sales representative for the most current information. Reagent Packs from the Gripper The reproduction, transmission or distribution Of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data patients, parameters and configuration dependent designations are fictional and examples only. HeoLtnineers All rights, including rights created by patent grant or registration of a utility model or design, are reserved. on the AteLLica@ 1M Analyzer connected to pcc Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands ore the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. 11 Copyright O Siemens Healthcare GmbH 2020 Patmt Sample Processtm Sample Processtm Global Siemens Healthineers Headquarters Henkestr. 127 Front Cover Re«mt voter 91052 Erlangen, Germany Phone: +49 913184 0 Back Cover siemens.com/heolthineers

  • atelica
  • attellica
  • attelica