MR Spectroscopy Peak Editor - Numaris X
This Job Aid provides workflow to add a Spectral peak to a theoretical Spectrum and save it as a customer protocol.
MR Spectroscopy Peak Editor Numaris X – Job Aid MR Spectroscopy Peak Editor – Numaris Job Aid If the Spectrum shows a peak that is not included in the theoretical Spectrum (e.g., Lactate peak), the peak can be added, the curve fit recalcuated and a customer protocol saved with the peak added. 1. Open Patient Browser 2. Select MR Spectroscopy (SVS or CSI) Dataset a. Right Mouse Assign > MR Spectro Workflow i. Workflow Assigned > Double-click on Patient > Opens in MR Spectro Workflow Patient Browser Local Data Search by Name, ID, Date, Description and Modality (Ctrl+F) Q More Filters Results (52 results, 10:11:52 PM) Patient Name Patient ID Date of Birth Study Date and ... Study Descripti ... Accession Nu ... Workflow Stu View&GO Dat ... 123456789 1/28/1991 1/28/2021 9: ... View&GO Dat ... MR_Cardiac Pl ... 56342626 9/16/1968 7/22/2014 10. Chest^Thorac ... MR_Neurology Switzerland_ ... 1/1/1960 12: ... 7/25/2012 10 Head Ae Stro ... MR Cardiac_F .. LV flow 5/19/1982 7/11/2011 8: ... Cardiac^Stan ... MR Spectro Germany_MAC 12/1/1983 11/13/2003 7 Head Av Do MR Sp ... MR Prostate ,... CS EDU SY 7/4/2019 12: ... 1/1 Open Ctrl+C MR_Spectro Siemens_MAC 1/1/1984 4/1 Open with MR_Cardiac Germany_MAC 1/1/1980 10/ View as Read-Only with MR_Oncology Siemens MAC 1/1/1947 10/ MR_Whole_S ... Siemens MAC 1/1/1965 6/2 Batch Open MR Neurology Switzerland 1/1/1934 3/9 Assign Workflow Favorites MR Neurology Switzerland 10/12/2018 4/21 Cancel Workflow Suggested MR_Tissue4D Poland MAC 1/1/1993 3/27 .......----- MR BreVis Belgium_MAC Search Patient with DICOM Q/R 1/1/1974 12: ... 9/27 MR Neurology MR_Oncology Siemens MAC 1/1/1947 6/1 Add to Demo List MR Oncology MR Oncology ... UK MAC 1/1/1965 9/2 Remove from Demo List MR Prostate * MR_Oncology ... UK MAC 1/1/1965 5/7/ Set Mark state MR Breast Spain_MAC 1/1/1977 7/28 MR Spectro * MR_Angiogra ... Germany_MAC 1/1/1961 Set Delete Protection state 11/6 MR Tissue4D SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Healthineers Job Aid Effective Date: 06/01/2021 | HOOD05162003193565 Job Aid |MR Spectroscopy Peak Editor – Numaris X Effective Date: 06/01/2021 | HOOD05162003193565 3. MR Spectro Analysis a. Select SVS or CSI Long TE Spectrum (270) i. Right-click on SVS or CSI Protocol > select Edit Protocol MR Spectro MR_Spec MR Spectro Analysis Germany *12/1/1983 csi se 135,11/13/2003 :22 PM CURREN Lu csi_se 135 Select Protocol ... csi_se 30,11/13/2003 3 Ly csi se 30 Edit Protocol ... svs_se 30,11/13/2003 3: Save Protocol svs se 30 svs_se_30,11/13/2003 3: Save Protocol As ... Ww SVs_se_30 Delete Export Export Export Spectro Compare CSI Selected Selected Selected Browser SP H47.6 SL 10 Tra Viewing O MPR THICK TR 9000 TE 100 .............. b. Protocol Editor Window i. Select Prior Knowledge Tab ii. Click [+] Protocol Editor - csi_se_135 Protocol name csi_se_135 Properties Prior Knowledge Post Processing Metabolite Water > Water > NAA Line shape Gaussian > Cr1 Use for match algorithm >Cho Tolerance 15 Hz > Cr2 Common Parameters Parameterization with model signal > Extra Parameters Model Signal File Siemens\1.5 T\1H\Water_se_135 Browse .. Spectral filter Note: This metabolite is used for reference processing. OK Cancel Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 06/01/2021 | HOOD05162003193565 Healthineers Page 2 of 8 Job Aid |MR Spectroscopy Peak Editor – Numaris X Effective Date: 06/01/2021 | HOOD05162003193565 c. Select Metabolite Template Window i. Select Metabolite Peak: [e.g., Lac_se_135] • Hint – match with protocol type selected sequence (se_135; st_135) ii. Click Select Select Metabolite Template ? X Show V W Siemens M In vivo Mi Customer Phantom View Details Systems Protocols Metabolite Templates Model Signal Files + 1.5T Metabolite Template 1H L Gln st 20 L Glu se 30 L Glu st 20 L Glx se 30 L Glx st 20 Le Lac_qa_se 135 Le Lac_qa_se_270 Le Lac_qa_se 30 Ly Lac_qa_st_135 Le Lac_qa_st 20 Ly Lac ga st 270 Lac_se 135 Le Lac_se 270 Le Lac se 30 Ly Lac_st 135 Le Lac_st 20 Le Lac st 270 Select Cancel d. Select Metabolite [e.g., Lac] i. Click OK Protocol Editor - csi_se_135 ? X Protocol name csi_se_135 Properties Prior Knowledge Post Processing Metabolite Lac_1 Water > Lac_1 Line shape Gaussian Scale Factor Frequency Use for match algorithm Width Phase Parameterization with model signal > Lac Model Signal File Siemens\1.5 T\1H\Lac_se_135 Browse .. > NAA Cri Spectral filter Pass band 0 ppm 3.9 ppm Cho > Cr2 Common Parameters OK Cance Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 06/01/2021 | HOOD05162003193565 Healthineers Page 3 of 8 Job Aid |MR Spectroscopy Peak Editor – Numaris X Effective Date: 06/01/2021 | HOOD05162003193565 e. New Protocol with Added Metabolite – Labeled with [*] i. Left-Click > Opens New Protocol with Added Metabolite shown in Spectrum MR Spectro Q + AH MR Spectro Analysis MR_Spectro Radiologie Uni Tuebingen Germany_MAC Ref.: Congress *12/1/1983, M, 19Y Avanto syngo MR 2004V 4VB11A csi_se_30,11/13/2003 3:32 PM CURRENT HFS t2_tirm_tra_dark-fluid +LPH csi_se_30 latest Spectro1.Reflmage.1 svs se 30,11/13/2003 3:37 PM STUDY 1aaaa1250 Wy svs se 30 11/1 3/2003 svs_se_30,11/13/2003 3:43 PM 3:00:54 PM Series 2 svs_se 30 R csi_se_135,11/13/2003 3:22 PM 1 cm csi se 135 1* SP H47.6 F SL 10 Tra MPR THICK MR MRI 2 IMA 1 Export Export Export Spectro Compare TR 9000 SP H37.1 Selected Selected Selected Browser TE 109 FoV 160*118 NA 1 Tra>Cor(-6) HE1-4 "tir 2d1_21/150 VOX AP LR FH SIZE 10 10 10 Viewing Voxel 106 I(NAA)/I(Cr1) 0.66 3.08 % I(Cho)/I(Cr1) 1.85 8.59 % MR_Spectro Radiologie Uni Tuebingen Germany_MAC Ref.: Congress *12/1/1983, M, 19Y Avanto Cho syngo MR 2004V 4VB11A CURRENT HFS csi_se_135 I: 4.26 +LPH latest/csi_se_13501 Spectrum -- > Spectro1 .Spectrum.13 STUDY 1aaaa1 250.35 11/1 3/2003 2.3 3:22:37 PM 0.30 Series 14 0.25 0.20 :1.51 R 0.15 Real Part 0.10 0.05 0.00 TTTTT TR 1500 -0.05 E 135 NA 4 4.0 3.5 3.0 2.5 2.0 1.5 1.0 0.5 TA 07:12 Chemical shift [ppm] VOX AP LR FH HE1-4_c_4 POS A38 L16 H41 csi_se/90 I: Integral SIZE 10 10 10 77 Voxel 106 Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 06/01/2021 | HOOD05162003193565 Healthineers Page 4 of 8 Job Aid |MR Spectroscopy Peak Editor – Numaris X Effective Date: 06/01/2021 | HOOD05162003193565 4. Result Table – Add Metabolite a. Select Upper Right Corner Menu > Define Result Table I(NAA)/I(Cr1) 0.66 08 % (Cho)/I(Cr1) 1.85 Define Result Table b. Define new value i. Backspace over current Entry up to [I(] ii. Enter I(Lac) iii. Click Add iv. Select OK Define Result Table ? X Define Result Table ? X Defined values Defined values I(NAA)/1 (Cr1) I(NAA)/I(Cr1) I(Cho)/I(Cr1) I(Cho)/I(Cr1) I (Lac) Define new value Define new value I(NAA)/I(Cr1) I(Lac Examples >> Add Examples >> Add OK Cancel OK Cancel Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 06/01/2021 | HOOD05162003193565 Healthineers Page 5 of 8 Job Aid |MR Spectroscopy Peak Editor – Numaris X Effective Date: 06/01/2021 | HOOD05162003193565 5. Save Customer Protocol a. MR Spectro Analysis a. Right-click New Protocol > Select Save Protocol As i. Enter Protocol Name ii. Deselect Show > SIemens iii. Click Save MR Spectro CE + MR Spectro Analysis OH csi_se 30,11/13/2003 3:32 PM Ly csi se 30 svs se 30,11/13/2003 3:37 PM Wy SVs se 30 svs_se_30,11/13/2003 3:43 PM Wy Svs se 30 csi se 135.11/13/2003 3:22 PM csi se Select Protocol ... Edit Protocol ... Export Save Protocol bare Selected Save Protocol As ... Viewing Delete Save Protocol As ... ? X Show [ Siemens In vivo View Details i Customer Phantom Systems Protocols Metabolite Templates Model Signal Files › 1.5 T 1H Protocol System Nucleus Sequence Purpose TE [ ms] Protocol name csi_se_135_Lac Save Cance Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 06/01/2021 | HOOD05162003193565 Healthineers Page 6 of 8 Job Aid |MR Spectroscopy Peak Editor – Numaris X Effective Date: 06/01/2021 | HOOD05162003193565 6. Open Spectroscopy Saved Customer Protocols a. MR Spectro Analysis I. Left-click > Spectro Browser II. Select Show > Customer Protocol i. New Customer Protocols shown in MR Spectro Analysis MR Spectro + MR Spectro Analysis OH csi_se_30,11/13/2003 3:32 PM Ww csi_se_30 i svs se 30,11/13/2003 3:37 PM svs_se 30 svs se 30,11/13/2003 3:43 PM Wy SVs se 30 iv csi_se 135,11/13/2003 3:22 PM csi se 135 Lac AB Export Export Spectro Browser Fompare Selected Selected Selected Viewing Select Protocol ? Show Siemens In Vivo i Customer View Details Phantom Systems Protocols Metabolite Templates Model Signal Files + 1.5 T 1H Protocol System Nucleus Sequence Purpose TE [ms] 31P ¡ csi_se_135_Lac 1.5 ₸ 1H csi_se n vivo 125-145 Select Cance Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 06/01/2021 | HOOD05162003193565 Healthineers Page 7 of 8 Job Aid |MR Spectroscopy Peak Editor – Numaris X Effective Date: 06/01/2021 | HOOD05162003193565 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com ___________________________ Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 06/01/2021 | HOOD05162003193565 Healthineers Page 8 of 8
- mr
- spectroscopy
- peak
- editor
- numaris x