MR Spine Autoalign Verification (E-line) Job Aid
MR Spine Autoalign Verification (E-line) Job Aid
MR Spine AutoAlign Verifica- tion (E-line) 1. Check for correct positioning and labeling of the vertebras in the reference images. a. To check positioning, scroll through the image stacks with the Series +/- and the Image keys. 301/1214 Dhot X Nitlige Veelcation Check te xxxt Contre Decine Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Job Aid | MR Spine Autoalign Verification(E-line) Effective Date: 05/04/2020 | HOOD05162003083148 2. Edit Vertebral Labeling a. Select the desired spine region (i.e., C, T, or L-spine) from vertebra selector graphic on the guidance card. AutoAlign Verification sbc + Show intervertebral disk positions mage Labeling Label transverse images Check the shown vertebra labeling and confirm below. Confirm Decline 3. Select the individual vertebra label from the “zoomed“ Vertebra selector graphic of the Guidance card that needs to be edited. a. On the Guidance card this icon is activated. AutoAlign Verification T3 V Show intervertebral disk positions Image Labeling Label transverse images Check the shown vertebra labeling and confirm below. Confirm Decline Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 2 of 5 Job Aid | MR Spine Autoalign Verification(E-line) Effective Date: 05/04/2020 | HOOD05162003083148 4. Click the wrong vertebra label in the reference image a. Automatic relabeling is performed on every correction SIM HR SIM SIM SIM 51 IMA 8 /15 10/19.2018 10/19/2018 C2 T4 75 10am PZ INDRI.(.DIS2D/F X3_2.FIL TP F211 SP 12.1 FoV34734) P3 CPRIORNIDIS3D/COLIP SP 18.2 12 tro x48 448_p2 Sag>Tin-t 4) I AASpine Scout COMP SP 1 FoV 800 183 Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 3 of 5 Job Aid | MR Spine Autoalign Verification(E-line) Effective Date: 05/04/2020 | HOOD05162003083148 5. Confirm the correction a. If Label transverse images is not selected, Confirm and Delete will not be selectable AutoAlign Verification abc Show intervertebral disk positions Image Labeling Label transverse images Vertebra labeling is confirmed. Transverse images will be labeled. Confirm Decline Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 4 of 5 Job Aid | MR Spine Autoalign Verification(E-line) Effective Date: 05/04/2020 | HOOD05162003083148 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 5 of 5
- Spine Autoalign Verification
- E-line