Performing IMT Advanced Clean on the Atellica® CH Analyzer Job Aid Version 1.25 and Below
This job aid provides instructions on how to perform the as needed maintenance task of IMT Advanced Clean for the Atellica® CH Analyzer.
Atellica® CH Analyzer SIEMENS Performing IMT Advanced Clean Job Aid Healthineers An alert generates for the operator to perform the IMT Healthingers Atellica Solution Version 1.14.0.2983003 SP00107 Advanced Clean after the analyzer processes 5000 IMT 160 System Samples Inventory Calibration QC Patient Orders Worklist 1 Maintenance Setup samples or every 30 days, whichever comes first. Schedule Report Log 2 IMT Data Restore Software Updates Copy Configuration CH After the analyzer completes the IMT Advanced Clean, CH 3 IMT Clean and Prime IMT Details IMT Messages the operator must replace the A-LYTE multisensor. IMT Status IMT Calibration completed 1. Select the appropriate IMT function. Prime IMT IMT Dilution Check MT Calibration 2. Select Prime fluids one at a time by selecting a fluid. IMT Routine Clean 1. At the Command bar, select Maintenance Last Performed 06/20/2018 9:01:08 PM Next Scheduled 06/21/2018 1:01:08 AM Operation Status 4 IMT Advanced Clean A-LYTE Multisensor 2. Select the IMT tab Prime All Date Last Replaced 05/29/2018 1:41:45 PM Tests on Chip 0 Prime Standard A 3. Select IMT Clean and Prime Routine Clean Prime Standard B Last Performed 06/20/2018 2:05:54 AM 4. Select IMT Advanced Clean Advanced Clean Prime Salt Bridge Prime IMT Diluent 5. Select OK to acknowledge that multisensor must be Last Performed Not Performed Days to next clean Maximum O replaced after IMT Advanced Clean completes Serum or Plasma samples processed Limit O Average Fluid Delta Not Available Minimum = 0.0 mv Sample Port Drain Time Not Performed Maximum = 1.7 seconds 6. Wait 14 minutes for advanced clean to complete Cance 7. Replace the current A-LYTE® multisensor ? PEP LabManager Thursday, June 28, 2018 12:36:26 PM connect Note: Do not perform an IMT Advanced Clean after replacing the A-LYTE multisensor. Performing the IMT Advanced Clean on the current multisensor renders it unusable. Perform the IMT Advanced Clean on the current multisensor prior to replacement. 1 Template Effective Date: 05/27/2019 | HOOD05162003010284 Effective Date: 24-SEP-2020 | HOOD05162003126734 Unrestricted © Siemens Healthcare GmbH, 2020 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2020 Global Siemens Healthineers Headquarters Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84 0 siemens.com/healthineers 2 Template Effective Date: 05/27/2019 | HOOD05162003010284 Effective Date: 24-SEP-2020 | HOOD05162003126734 Unrestricted © Siemens Healthcare GmbH, 2020
- attelica
- attellica
- atelica