Performing IMT Dilution Check on the Atellica® CH Analyzer Video Version 1.25 and Below

This video provides instructions on how to perform the as needed maintenance task of IMT Dilution Check for the Atellica® CH Analyzer.

An IMG dilution check can be performed on the Atellica solution Ch Analyzer as part of as needed maintenance in order to complete this procedure. First, stop the analyzer and unlock the front cover. This can be done by selecting system in the command bar. Next select status. Select the Ch icon. Select the stop button. Finally, select the button to unlock the front cover of the analyzer. Open the front cover of the Ch analyzer. The cover will automatically lock within 30 seconds of selecting the unlock button in the software. If the cover is not open within 30 seconds, the operator will need to select the unlocked front cover button in the software again. Fil A1 milliliter tube top sample Cup with 500 to 750 microliters of IMT dilution check. Place the tube top sample Cup in the dilution check position behind the a light multisensor. Close the cover and select the resume button on the Ch System status screen. On the Command bar, select maintenance. Next select IMT. Select IMT dilution check. Select perform dilution check. Finally, select OK. Wait for the empty dilution check to complete. When dilution check is complete, select apply correction in the software. Next, so like system in the command bar. Select status select the Ch icon. Select the stop button. Select the button to unlock the front cover. Open the front cover of the Ch analyzer. Remove and dispose of the dilution check Sample Cup according to laboratory protocol. Close the cover and select the resume button in the software. To complete the procedure, review the report for acceptance criteria by searching for AMT dilution check in the Atellica solution online help.

Atellicae AtelLicae Solution AtelLicae Atellicae Solution SIEMENS . SIEMENS Solution Version 1. Solution sp00107 Version 1. 14.0.2983003 Heo[tnineers HeoLtnineers HeoLthineers SIEMENS . HeoLtnineers Healthineers I ry Operator Diagnostics Oget-atod.Z@gnostics D Restore Restore Software Updates Software updates Copy C Copy Hdp Help SH CH 1M C.F1-05 System VMM C-P1.05 Dilution Check status Dilution Check Dilution Perform Check 1M T cwan ana pnme 1M T clean ana pnme 1M T clean anc pnme 1M T cwean ana pnme Current Factor Unknown 'MT Dilution Check date 'MT Dilution Check completed at 1110512018 Standby Standby 'MT mutton Check 1M T mutton Check 'MT o.ution Check mutton Check Check Apply Correction 140.78 141.28 140,7 CH 400 DilCheck Search Library Capabilities the correction About IMT Dilution Check The is the on the Chock is within Maintenance 14162 Dilution Check Dilution Check Connected to FCC 389 K means. In to 138.2 1369 3.978 The is a IS a 140 7 Front Cover Front is the Na and mean Na and the Na and mean the Na 137 s '40.9 e •s car Back Cover Front Cover Cover 13745 137.45 141.68 3.883 3.978 3.978 Bottle Value 4.0 %Bias 40 PEP LabManager Monday, 05, 2018 Tuesday, AM Tuesday. July 10.2018 Tuesday, July 10.2018

  • attelica
  • attellica
  • atelica