PEPconnect

Processing Hemoglobin A1c Samples Job Aid

This job aid provides instructions on how to process Hemoglobin A1c samples on your DCA Vantage® Analyzer, including preparing the cartridge with the sample, inserting the test cartridge into the system and removing it when the test is completed.

HbA1c Sample Processing Performing a Hemoglobin A1c Test Running Quality Control 1. Touch the tip of the capillary to the blood 8. Remove and discard the cartridge sample until the capillary is filled • Select Next, then open the door 2. Wipe the side of the capillary tube with a lint- • Press and hold the plastic tab with the left hand free tissue • Gently push down and to the right to unlock • If an air bubble(s) is present in the capillary tube, discard and repeat 3. Insert the capillary holder into the cartridge until it snaps into place 4. Scan the cartridge DCA Vantage • Barcode facing to the right SIEMENS • Insert into barcode track above the dot • Slide cartridge smoothly down track • Audible beep sounds when successful 5. Insert the cartridge • Barcode facing to the right SIEMENS DCA" Systems 0975 • Insert cartridge until a click is heard LOT • Slowly and firmly pull flexible tab 6. Close the compartment door 7. Record the results GPE15009.001 Effective date 06/09/2014

  • sample
  • processing
  • hemoglobin
  • A1c
  • HbA1c
  • HgbA1c
  • cartridge
  • barcode
  • capillary
  • blood
  • job aid