Processing Microalbumin/Creatinine Samples Job Aid
This job aid provides instructions on how to process Microalbumin/Creatinine samples on your DCA Vantage® Analyzer, including preparing the cartridge with the sample, inserting the test cartridge into the system and removing it when the test is completed.
Microalbumin/Creatinine Sample Processing Running a Microalbumin/Creatinine Test Running Quality Control 1. Insert the tip of the capillary tube into a room 8. Close the compartment door temperature, well mixed urine sample to a 9. Record the results level just above the starch plug 10. 2. Wipe the side of the capillary tube with a lint- Remove and discard the cartridge • Press and hold the plastic tab with the left hand free tissue • If an air bubble(s) is present in the capillary Gently push down and to the right to unlock • tube, discard and repeat 3. Insert the capillary holder into the cartridge until it snaps into place 4. Scan the cartridge • Barcode facing to the right • Insert into barcode track above the dot • Slide cartridge smoothly down track • Audible beep sounds when successful 5. Insert the cartridge • Barcode facing to the right • Insert cartridge until a click is heard 6. Depress the plunger into the cartridge LOT 0073 € 7. Slowly and firmly remove the flexible pull tab ams GPE15009.001 Effective date 06/09/2014