Reviewing QC Job Aid
This job aid provides instructions on how to review Quality Control on your RAPIDComm® Data Management System, including how to electronically record the review of multiple QC results.
Reviewing QC Reviewing QC results 1. Navigate to Reports > QC. 2. Expand the tree in the left pane and select the analyzer. 3. Use the filters specify the date range, QC levels and types of QC results you wish to review (accepted, rejected, discarded) and review status (all, reviewed, not reviewed). 4. Use the Samples, Statistics and Charts tabs to review the QC data. 5. To modify a QC result record, select the result record on the Samples tab and select View… 6. To add comments, change the Operator ID or reclassify a QC result, select Edit. 7. After making the desired changes, select OK. 8. To electronically record the review of multiple QC result, select the first QC result record at the top of the Sample table, scroll to the bottom of the table, hold down the shift key and select the last result record. 9. Select the Record Review button and then select OK. 10. For each of the result records the Reviewed column is updated with the date and time and the Reviewer column with your user name. GPE11001.008 Effective Date: 07/23/2014