Recovering from Sample Travel Overage on the Atellica® Solution Job Aid Version 1.25 and Below
This job aid provides details on the topic of recovering from sample travel overage on the Atellica® Solution.
Atellica® Sample Handler Connect SIEMENS Healthineers Recovering from Sample Travel Overage Job Aid Recovering from Sample Travel Overage Sample Travel Overage occurs when there is a condition that prevents samples from unloading from the Atellica Magline® Transport, such as the Laboratory Automation System (LAS) or SHC being unavailable to unload LAS samples, and the SH also being unavailable to unload samples. Under this condition, the samples continue to travel on the Atellica Magline Transport until the condition is resolved or the distance the sample travels approaches the Sample Travel Overage limit. When the sample travel approaches the Sample Travel Overage limit: • The system alerts the operator and pauses the Atellica Magline Transport • If the operator resumes the Atellica Magline Transport, the system attempts to unload the sample to an available rack in an SH drawer. If the system still cannot unload the sample, the system alerts the operator and pauses the Atellica Magline Transport again • Each time the operator resumes the Atellica Magline Transport, the sample moves farther along the Atellica Magline Transport until the system successfully unloads the sample or the Sample Travel Overage limit is reached When a sample meets the Sample Travel Overage limit: • The system alerts the operator and pauses the Atellica Magline Transport • The system will not allow the operator to resume until the operator reseats all samples in the carriers currently on the Atellica Magline Transport and restarts the Atellica Magline Transport Template Effective Date: 27-MAY-2019 | HOOD05162003010284 1 Effective Date: 05-JUN-2019 | HOOD05162003006834 Unrestricted © Siemens Healthcare GmbH, 2019 Atellica® Sample Handler Connect SIEMENS Healthineers Recovering from Sample Travel Overage Job Aid To recover from Sample Travel Overage: SIEMENS Healthineers Atellica® Solution Version 1.16.0.3124001 SP00101 A 1. Ensure the SH has not stopped or paused due to an error System Samples Inventory Calibration Patient Orders Worklist Maintenance Setup Status Logs Operator Diagnostics About condition: System VMM P-P1-04 SC-P1-03 C-P1-05 P1-02 CM00301 A. On the Command bar, select System > Status B Standby Standby Standby Standby Standby Standby B. Select the SH icon Diagram Alerts Capabilities C. If the SH stops or pauses due to an error condition, correct Carriers Atellica Magline Transport the error condition and restart the SH if appropriate 2. Load an appropriate rack with available positions in an SH Connected to PCC SC-P1-03 P-P1-04 C-P1-05 I-P1-02 CM00301 drawer 3. To unload samples from the Atellica Magline® Transport, start the Atellica Magline Transport: A. On the Command bar, select System > Status B. Select the Atellica Magline Transport icon ? PEP. LabManager Tuesday, November 13, 2018 12:12 PM C. Select the Resume icon *This procedure is continued on the next slide 2 Effective Date: 05-JUN-2019 | HOOD05162003006834 Unrestricted © Siemens Healthcare GmbH, 2019 Atellica® Sample Handler Connect SIEMENS Healthineers Recovering from Sample Travel Overage Job Aid 4. If the system does not restart the Atellica Magline® Transport due to the Sample Travel Overage limit being met or exceeded, perform the following: A. Visually inspect the Atellica Magline Transport to locate any carriers with sample containers B. To gain access to the carrier, remove 1 or more Atellica Magline Transport covers: 1) Loosen the 2 Atellica Magline Transport cover thumbscrews enough to allow you to slide the cover off, but do not completely remove the thumbscrews 2) Pull the Atellica Magline Transport cover away from the system C. Push the sample container down until it is fully seated to the bottom of the carrier 001 1001234 5997 D. Ensure carriers are clear of the Atellica Magline Transport diverters. The Atellica Magline Transport and SH do not initialize if a carrier is present in the diverters E. If appropriate, ensure all analyzer probes and SH / SHC grippers are off the Atellica Magline Transport and in a safe position before starting the system. This enables the free movement 210914552A-211 of the carriers when the Atellica Magline Transport initializes Carrier at diverter position *This procedure is continued on the next slide 3 Effective Date: 05-JUN-2019 | HOOD05162003006834 Unrestricted © Siemens Healthcare GmbH, 2019 Atellica® Sample Handler Connect SIEMENS Healthineers Recovering from Sample Travel Overage Job Aid F. Replace the Atellica Magline Transport cover(s): 1) Position each cover so the edges of the cover fit into the slots on either side of the opening 2) Slide the Atellica Magline Transport cover in until the spring clips lock the cover into place 3) Tighten the 2 thumbscrews under the Atellica Magline Transport cover G. Start the Atellica Magline Transport by selecting the Resume icon on the Atellica Magline Transport System Status screen 4 Effective Date: 05-JUN-2019 | HOOD05162003006834 Unrestricted © Siemens Healthcare GmbH, 2019 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2019 Global Siemens Healthineers Headquarters Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84 0 siemens.com/healthineers 5 Effective Date: 05-JUN-2019 | HOOD05162003006834 Unrestricted © Siemens Healthcare GmbH, 2019
- atelica
- attellica
- attelica