Replacing the Dryer Nozzle Probes on the Atellica® CH Analyzer Video Version 1.25 and Below
This video provides instructions on how to perform the as needed maintenance task of replacing the dryer nozzle probes on the Atellica® CH Analyzer.
show you the steps on how to replace the Atellica Chemistry analyzer dryer nozzle probes in order to change the dryer nozzle probes, the system must be set to diagnostic state. This can be done by selecting system in the command bar. Next select operator diagnostics. Select the chemistry module icon. Select enter diagnostics. Select yes in subsystems, select probes. Select dilution. Wash probe dryer one or reaction wash probe dryer one select perform. Select yes, wait for the analyzer to prepare for maintenance. The covers automatically unlock if the back cover re locks before you were able to open it, select the button in the software to unlock the back cover. Open the chemistry analyzer back cover. There is one dryer nozzle probe for the reaction ring and another dryer nozzle probe for the dilution ring. The two dryer nozzle probes are interchangeable. Before starting to remove one of the dryer nozzle probes, cover the cue vets with a paper towel to protect them from debris. Unscrew the probe housing assembly thumbscrew counterclockwise. To access the dryer wash probe, lift the probe housing up and turn it over. Locate the appropriate dryer nozzle probe and disconnect the tubing. Pull the lever back to release the probe from the probe housing. Dispose of the probe spring an nozzle per laboratory protocol. Locate the new dryer nozzle probe and spring insert nozzle onto probe until it clicks. Insert the spring onto the probe. Insert the assembled probe into the probe housing until it clicks. Turn probes housing over an reinserted on the analyzer, tighten thumbscrew clockwise and connect the tubing 5 to 6 millimeters onto the probe. Remove the paper towel. Enclosed the back cover. Select continue. Select close, select exit diagnostics. Select yes select the maintenance tab in the command bar, select schedule, select the chemistry icon in the dropdown menu. Select the correct dryer nozzle probe. Select the mark as completed button on the right hand side of the screen to complete the procedure. Perform Ch auto check in operator diagnostics.
Atellicae Solution Version 1.14.0.2983003 Atellicae Solution Version 1 Atellicae Solution Version I Atellicae Solution Version I .14.0.2983003 Atellicde Solution Version 1.14.0.2983003 Atellicae Solution Version 1. SIEMENS SIEMENS . SIEMENS .. SP00107 SP00107• .14.02983003 .14.0.2983003 Healthineers SIEMENS .. Heolthineers Healthineers • System Subsystems M aimenarx:e Seno Setto Invaltory Inve•tory Invertory Inveritory Inventory invalory Invert*ory Category Orders OrdUrs Mültenarx:e Report Repert Exit Diagnostics Operator Operator Diagnostics CH 1M CH ATELLICACH ATELLICA.CH ATE-LUICA.CH Module: ATELLICA-CH Frequency Al Tune: All Performed By: All Module ATELLICACH Frequency All Time Al Performed By All System VMM ATE LUCA ATELLICA.CH Sun-27 Mon-28 Module Activity Fri-25 Sat-26 sat-Zb Subsystems Diagnostic Name Diagnostvc Name Descnption Descniption Diagnos c Nam Name Description Frequency Replacing Reagent Mixer Impeller Replacing Reagent or Wesh Replacing Searenple Mixer Impeller As Needed Standby Confirmation C O f i Reaction Wash Probe 2 Standby the Reaction Wash Probe Dryer 1 ing the Reaction Wash Probe 3 Reætion Wash Pro& 2 Rep-lacing the Reaction Wash Prober 2ozzle Replacing the Reaction Wash Probe 2 SH Probe CH Autocheck CH Autocheck Perform Performed All Time Rep{acing the Reaction Wash Probe 2ryer.ATELLICA Replacinvg Dilution Wash Probe 2 Replacing Sample Impeller Not Perform... Perform Replacing the Reaction Wash Prober 6ozzle Replacing Reaction Wash Probe 3 Reaction Wash Probe Dryer Dilution Wash Probe 3 Rextön "'Mash Probe 3 Replacin.g the Reaction Wash Prober 6ryzler Replacir•vg Dilution Wash Probe 2 Replacing the Reaction Wash 6 the Reaction Wash Probe 5 Category Tme Confimation All Capabilities Replacing Dilution Replacing the Dilution Wxer Impeller Replacing the Dilution Wash Probe Dryer 1 Needed the Procedure FWacim the Reaction Wash Probe 4 the Reaction Wash Probe 4 Rheacilution Wash Probe Dryer the Diaction Wash Probe 4 Reætim Wash Probe 4 Replacirothe Reaction Wash Probe 4 Replacim Reaction Wash Probe 4 IMT Replacing the Reaction Wash Probe 6 Replacing the Dryer Nozzle Probes Enter the activity narne. Search Text Replacing the Dilut•on Probe Replacing the Reaction Wash Prober 6ozzle Rep{acing the Reaction Wash Probe 2 Replacen.•g the Reaction Wash Probe Dryer: Replacin,g the Reaction Wash Probe 2 Replacing the {Reaction Wash Probe 2 Re-pltacin.g the Reaction Wash Probe 2ry'er 1 the Reaction Wash Probe 4 Replacing the Dilution Wash Probe Dryer: Replacing mhe Reaction WVash Probe Oryer ATELLICA Schedubed by Operator Scheduled by Operator Replacing Rheacilution Wash Dryer Nozzle Replacing the Wash Dryer Nozzle Perform: Replacing the Reaction Wash Probe Dryer? Perform Replacing the Reaction Wash Probe Dryer? Dilution Wash Probe 3 Replacing Reaction Wash 6 Completing Replacing the Reaction Wash Probe Dryer Schedubed Automatically Replacing the Dilution Wash Prober 1ozzle Replacing the Reaction Wash Dryber Nozzle Replacing Reaction Wash 1 Replacir.g Dilution Wash Probe 1 As Needed Autocheck Replacirvg Dilution Wash Probe 2 Replacinvg Dilution Wash Probe 1 Rep•tacir.g the Dilution Wash Probe 2 Replacing Dilution Wash Probe 1ryer Replacing the Dilution Wash Probe Dryer 1 Reptaeng the I.Vash Probe Replacing the Dilution Wash Probe 2 Rro the Dilution Wash Probe 1 Ing the Dilution Wash Probe 1 the Dilu%on Wash Probe probe Scheduled by Service Replacing Wash Probe 2 Needed As Needed Dilution Wash Probe 2 Connected to PCC Replacirvg Dilution Wash Probe 2 Replacir•.ghe Dilution Wash Probe 2ryer Rep-lacin.g Dilution Wash Nrozzle Reßacing Dilution Wash Probe 2 Yes No Replacing the Dilution -Wash Probe Dryer 1 Replacing the Reaction Wash Prober 3ryzler the Wash Probe 3 Replacing the Dilution Wash Nryber Nozzle Replacing Dilution Washe 2 Dilution Wash Probe 3 Inrvspect and Clean Inspect and Clean Replacing Dilution Dilution Wash Probe 3 Reaction Wash Probe the Reaction Wash Probe 5 Replacing the Dilution Wash P Rep'ac.nq the Wash Probe a the Dilution Wash Probe 3ryer Rheacilution Wash Probe Dryer Inprocess Replacng Reaction Wash ProbC4, Replacing the Reaction Wash Probe 1 Replacing Dilution Wash Probe 2 Replacing the Wash Probe Dryer I Reaction Wash Probe Dryer 1 Continue Close Yes No Replacir.g the Wash Probe Dryer Wash Replacing Reaction Wash Probe Dryer the Diaction Wash Probe 4 Reaction Wash Probe Dry Completed Front Cover Back Cover Replacin.g Dilution Wash Probe Dryer 1 Reaction Wash Probe Dryer the Dilution Wash Probe 3ryer Rheacilution Wash Probe Dryer the Dilution Wash Probe Dryer Rep-lacing the Dilution Wash Probe Dryer 1 Replacinvg the Dilution Wash Probe 1ryer 1 Replacing the Dilution Wash Probe Dryer 1 Replacir,g the Dilution Wash Probe Dryer 1 Re-PAaciro Dilution Wash Probe Dryer 1 the Dilution Wash Probe Dryer 1 Rheacilution Wash Probe Dryer 1 the Not Performed Not Performed_ Performed the Reaction Wash Probe 6 Reaction Wash Probe Dryer Dilution Wash Probe 3 Reaction Wash Probe Needed Back Cover Replacir.g the Reaction W'lash Dryer Nozzle Replacing Wash Dryer Nozzle Rep-lacimg the Reaction Wash Dryer Nozzle the Reaction Wash Dryer Nozzle the Dilution Wash Probe 3ryer Replacing the Dilution Wash Dryer Nozzle the Reaction Wash Dryer Nozzle Replacim the Reaction Wash Dryer Nozzle he Reaction Wash Dryer r-bzzle Reaction Wash Dryer NOzzie- With Comments Replacing Dilution Wash Probe 3 Replacing the Dilution Wash Probe Oryer 1 Replacing Dilution Wash Oryer Nozzle Replacin.gie Dilution Wash Dryer Nozzle Rep-lacimg the Reaction Wash Dryer Nozzle Rep-lacing the Dilution Wash Dryer Nozzle Replacimg the Reaction Wash Dryer Nozzle Rheplacimg Rheacilution Wash Dryer Nozzle Replacim the Reaction Wash Dryer Nozzle Rep-lacing Dilution Wash Nozzle Rep/acing Dilution Wash Dryer Nozzle Replacing the Reaction Wash Prober 2ozzle Replacing Dilution Replacing Rheacilution Wv'v'ash Dryer Nozzle Replactng the Dilution Nozzle Dilution Wash Dtyer Nozzle Nozzle Wash ATELLICA-CH Replacing Reagent Probe 1 Replacing Reaction Wash Probe 1ryer 1 Replacing the Reacect: Probe 3 Needed Exiter Diagnostics Operator Diagnostics Exiter Diagnostics Exitr Diagnostics Operator Diagnostics Exit Diagnostics Add Activay _ Set Warning. Set Warnirv.. PEP _PEP LabManager Monday, January 28, 2019 20724 PM Monday. January 28. 2019 247249 PM Monday, January 28, 2019 2424 PM Monday, January 28, 2019 2024 PM LabManaøer Monday, January 28, 2019 207249 PM Monday. January 28, 2019 Tuesday, July 10. 2018 112730 AM Monday, January 28, 2019 24249 PM January 28.2019 24249 PM LabManage
- attelica
- attellica
- atelica