PEPconnect

Respiratory Gating: Reconstruct Image with 10 Time-Based Phases

Respiratory Gating: Reconstruct Image with 10 Time-Based Phases step by step

Respiratory Gating: Reconstruct Image with 10 Time-Based Phases Computed Tomography  Phases are calculated based on fractions of the total length of the breathing cycle (10% = Patient completed 10% of one breathing cycle) 1. Register patient. 2. In the Patient Model Dialogue Box, select the desired respiratory gating protocol from within the RT protocol button 3. Left-click on the Trigger sub-task card to select it. 20 40 50 60 70 80 90 100 Original RESP < Average CT III uto Tasking Trigger 4. Left-click to select the Multiphase reconstruction job (recon job 1) RT_Resp (Adult) Topogram A, Resp RT > 12 rpm 5. From the drop-down on the Trigger sub- Increment 20 % In/Ex task card, select "Increment 10%" Increment 10 % In/Ex Increment 5 % In/Ex Increment 1 % In/Ex Increment 10 % Increment 5 % Increment 1 % Increment 10 % Pi Increment 5 % Pi Increment 1 % Pi 10 20 Clear selection SIEMENS siemens.com/healthineers Healthineers . Job Aid | Respiratory Gating: Reconstruct Image with 10 Time-Based Phases Effective Date: 12/6/2016 | Quality 1406 6. Left-click on the "0" on the left end of the bar and drag to "90" on the right to select multiple phases ("0" and "100" are the same phase). 0 10 30 40 50 60 70 80 90 100 0 10 20 30 40 50 60 70 80 90 100 Selecten1220% 40% 50% 60º. 70% 80% Original RESP 90% Average CT Average CT III Auto Tasking Trigger econ Auto Tasking Trigger Siemens Healthineers Services Page 2 of 3 Job Aid | Respiratory Gating: Reconstruct Image with 10 Time-Based Phases Effective Date: 12/6/2016 | Quality 1406 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright @ Siemens Healthcare GmbH ...... ........... Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 3 of 3