Scanning MC TDEF Barcodes on the Atellica® IM Analyzer Video Version 1.25 and Below
This video provides instructions on how to scan MC TDEF barcodes on the Atellica® IM Analyzer. For optimal viewing, select the arrows on the right of the playbar after the video begins.
The master curve and test definition 2D barcodes or MCTF bar codes must be scanned whenever a new lot of reagent is loaded on an atellica immunoassay analyzer. The Emct Def bar codes can be found in the immuno Assay reagent kit box. To scan the MCT DEF barcodes, first check the system status screen for the module to ensure the analyzer is in one of the following states. Ready standby stopped startup or warming up. Then at the Command Bar select set up. The first tab is test definition. Select I am test definition. Select the scan barcodes button. Scan each of the EM CTEPH barcodes in consecutive order. There may be one or more bar codes to scan. The barcode scanner will beep to confirm successful scans. To complete the procedure, select the Save button to save the scanned master curve and test definition information.
Atellicae Solution Version 1.14.0.2951003 SIEMENS . SIEMENS Solution Version 1.14.0.2951003 Atellica p_pl -03 HeoLtnineers Healthineers In Tesult 1M Test Definition Test Definition Settings Operator OiMostics Bastem System Wastem VMM SPOOIII CMOOIIO 00104 Assay The manufacturer encodes a collection ot data on a set of MC TDEF 2D barcodes C H Test CH Definition that are specific to an individual assay reagent lot and test definition version. Reselt Result Teselt The system uses the encoded values to calibrate, process, and calculate results. CH Definition 1M Wash Processing Group Test Information Ratios. and Otf• Test Type Repeat Conditions. Reøeat Conditiorw Capabilities The test demition scan was successful for the follcwing: stem System 10 Test Definition Test Version Test version BNP Test Display Display SIEMENS SIEMENS . 15 10 The master curve scm for the following was succes
- attelica
- attellica
- atelica