Performing a Basic Test Job Aid
This job aid provides instructions on how to perform a basic test on your Stratus® CS Analyzer. It also includes a list of the general supplies needed to operate the analyzer.
Stratus® CS System Basic Operation Refer to the Stratus® CS Operator’s Guide for additional information. 1. Compare the tube to 2. Scan the barcode label 3. If necessary, manually 4. Invert the tube gently the sample volume on the tube. enter the information a few times. Insert gauge. Note: If barcodes are not using the touchscreen. the tube into the used, go to step 3. Press Enter and select cannula until it snaps Accept (F4). into place. Doer Tes SIEMENS 5. Insert the cannula/tube 6. Load a new rotor in 7. Load up to 4 TestPaks 8. Press the green Start assembly into the the rotor table. Make in the Pak Manager key. sample door opening certain that the small well. and close the door. opening faces up. © Siemens AG 2013. All rights reserved. GPE15008.026 Effective Date: 10/22/2013 Stratus® CS System Basic Supplies Refer to the Stratus® CS Operator’s Guide for additional information. Sample Cup (CCUPS) Cannula - BD Tubes: Clear (CCANBD) Waste Container (CLINER) Sarstedt Tubes: Green (CCANS) EBNP TestPak CTNI DilPak SIEMENS SIEMENS SNP CalPak CTXI TestPak SIEMENS SIEMENS X Rotor (CROTOR) Paks - (See individual product Tip Tray (CTIP) inserts for catalog numbers) © Siemens AG 2013. All rights reserved.
- Basic Operations
- Basic supplies
- Sample Cup
- Cannula
- Rotor
- Paks
- Tip Tray