Performing Instrument Daily Setup Job Aid
This job aid provides instructions on how to perform Daily Setup activities on your Stratus® CS Analyzer, including replacing the waste container, and verifying both the tip supply and the printer paper supply.
Daily Setup Perform the following procedures before processing patient samples. Replace the Waste Container Verify the Tip Supply 1. Open the left front door. 1. Open the tip manager door. 2. Carefully pull out the waste container. 2. Remove empty tip trays. 3. Place a lid on the used waste container. 3. Insert new, full tip trays. 4. Dispose of the used container according to facility 4. Close the tip manager door. specific procedures. 5. Insert a new waste container. 6. Close the left front door. GPE15008.029 Effective Date: 10/22/2013 Verify the Thermal Printer Paper Supply 1. Select the Paper Advance key. If the edges of the paper are pink, replace the paper. 2. Lift the printer lid; remove and discard the used paper roll from the spindle. 3. Insert the spindle into the new thermal paper roll. 4. Position the spindle in the holder so that the paper feeds from the bottom of the roll. 5. Thread the paper through the roller. 6. Select the Paper Advance key three times. 7. Feed the paper through the slot in the printer lid and close the lid. 8. Select the Paper Advance key until about 5 cm (2 inches) of paper are visible. GPE15008.029 Effective Date: 10/22/2013 Run a System Check 1. Select Change Mode (F3) from the Patient Mode screen. ............................ 2. Close the sample door. System Check Pass 3. Use the arrow keys to move the cursor to System Check Mode; select Accept (F4). osten gnment test PASS Value Limit 4. Enter the System Check ID; select the Enter key. Alignment Value 139 ystem tendersture test 5. Enter the Operator ID; select the Enter key. PASS Uslue Temperature Linit 37.0 36.8 - 37.2 6. Select the green Start key. Fluidic sv PASS 7. Review the printed report. Value Level 1 Limit If the status of all components is PASS, proceed 10 Lavel 2 press €34 with processing. Pressure setpoint 435 370 600 Pressure Pande 720 If the status of any component is FAIL, refer to the System fluorome PASS System Check Troubleshooting section of the Value Linit Operator’s Guide. Level 1 0.513 ¢ 200 Level 2 467.2 Elec: Noise 453.65 - 47 0. 141 ¢ 4.5 8. Retain System Check reports with your quality control records. Printed: @ /23/2013 13:24 GPE15008.029 Effective Date: 10/22/2013
- job aid
- daily setup
- replace waste container
- verify tip supply
- verify printer paper supply