Performing a Quality Control Test Job Aid
This job aid provides instructions on how to test Quality Control on your Stratus® CS Analyzer, including preparing the QC material and processing it on the instrument.
Quality Control Preparing QC Material Prepare the control material according to the manufacturer’s instructions. Thoroughly mix the Mode Method material before placing it in a sample cup. Patient System Check Mode CKMB QC Low Mode Processing QC Hi Mode Calibration Mode Miso Mode 1. Fill the sample cup to the bottom edge of the Manual Dilution Next Mode Method Accept rim. Ensure there are no particulates in the samples and no air bubbles on the surface. F1 F2 F3 F4 2. From the Patient Mode screen, select Change Mode (F3). ? Exit 3. Highlight the required QC Mode. 4. Select Next Method (F3) to select an assay. Sample Select Accept (F4). 7 9 Delete Door 5. Foto 6. Select the Enter key. o Cup 4 5 6 a Pak 7. If required, enter the Operator ID. @ Tips 8. Place the sample cup in the cup position on 1 2 3 Waste the rotor table. Enter 9. Close the sample door. o 10. Place one TestPak per assay in the Pak < > Manager well. 11. Select the green Start key. GPE15008.029 Effective Date: 10/22/2013
- job aid
- quality control
- QC
- prepare QC material
- prepare Quality Control material
- process QC
- process Quality Control