Testing Samples Job Aid
This job aid provides step-by-step instructions on how to use your RAPIDLab® 1200 System for sample testing, including how to review results and simple troubleshooting tips.
RAPIDLab® 1200 Series User Guide Answers for life. SIEMENS Step-by-Step 1. Enter or scan password if prompted. 5. Enter the demographics if prompted, followed by the green continue button. 2. Mix the sample thoroughly. 6. When prompted remove the sample device as shown on the screen and press the green continue button. 3. Check on the screen the sample device 7. View the results. When completed the you are using is highlighted, if not touch results will be printed. Press the green the device to change it. continue button when done. 4. Insert the sample device as shown on the screen and press the green analysis button. Troubleshooting Touch to view the video if you need help. 1. If there is an obstruction (possible clot in the sample), you will be required to change the sample port as shown on the screen. 2. To view a video of the procedure press the grey camera icon next to point 1. 3. When you have attached the new sample port, press the green button. Never use an old sample Replace the sample port. For more information, please visit Siemens at www.usa.siemens.com/bloodgas Order No. A91DX-0701736-UC1-4A00 | Printed in USA | © 2010 Siemens Healthcare Diagnostics Inc. | www.usa.siemens.com/diagnostics
- job aid
- sample testing
- troubleshooting
- step-by-step