Unloading All Cal-QC Tubes Manually Video Version 1.25 and Below
This video will explain how to unload all Cal-QC Tubes manually.
This procedure to manually unload all tubes from the Calcu, See storage area should only be performed if the system is unable to unload the calcu. See storage area using the robot, such as during a power failure or robot failure. If the system still has power, first put the sample handler into stop mode by selecting the system button on the command bar. And then selecting the status tab. And then a sample handler module icon on the sample handler status screen. Select the stop icon. Once the sample handler is stopped, open the sample handler back cover by first selecting the unlock back cover button. Then open the back cover. Manually move the SH robot to the back corner of the SH. Do not hold the SH robot grippers to move the robot. Only move the robot by holding the robot body. Ensure the SH robot and cabling to not block the calcu. See storage area cover. Loosen the two thumb screws on the back of the calcu. See storage area cover. Rotate, open the calc UC storage area cover. Then remove all calc tubes from the storage area. For each tube, the tube cover must first be removed, and then the tube can be removed after removing the tube, replace the tube cover and then move on to removing the next tube. When replacing the tube covers, ensure that the tabs on the tube cover align so the cover properly seats in the Cal QC storage area. After removing all calc tubes and reinstalling the tube covers, tube covers must be present on all positions except the three rightmost positions. If appropriate, clean the calcu. See storage area by first wiping the area with lint, free tissues or gauze saturated with a 10% bleach solution. Then wiping with lint free tissues or gauze saturated with water and finally drying the area thoroughly with clean lint, free tissues or gauze. Loans the calc you see storage area cover. Tighten the two calcu. See storage area cover thumbscrews, then closed the sample handler back cover. Once the sample handler back cover is closed, lock the cover my selecting the lock back cover button. Start the sample handler by selecting the resume button on the sample handler system status screen, and then complete the procedure by confirming that the Calcu See storage area contents have changed to the default. On the sample handler system status screen. Select the confirm Calcu see storage content button. And then select the all samples have been removed by the user and contents restored to default button followed by the OK Button.
SIEMENS Atellica• Solution Healthineers Samples Inventory Calbabon Calbation Calbabion PatieN O rckrs Patient Orders Patient O rckrs PatieN P atieN O rckrs Setup Seartup Startup Please note that the learning material is for training purposes only! operator Operator Diagnostics For the proper use Of the software, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") Diagnose— Diagnostics Diagnose—cs issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software System .01 CH IA-03 and hardware available at the time Of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Processing Standby Standby Stopped ry Note: Some functions shown in this material are optional and might not be part of your system. Tirne ry Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be comn mercially available in your country. Capabilities Due to regulatory requirements, the future availabhility of said Functionalities in any specific country is not guaranteed. Please contact your local SIEMENS Siemens Healthineers sales representative for the most current information. Robot The reproduction, transmission or distribution Of this training or its contents is not permitted without express written authority. Offenders will be Liable for damages. TCS All names and data Of patients, parameters and configuration dependent designations are fictional and examples only. Healthineers Cal-QC storage cal-ac storage All rights, including rights created by patent grant or registration of a utility model or design, are reserved. connected to pcc Front Cover Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Drawer 3 Back Cover product availability varies by country and is subject to varying regulatory requirements. please contact your local representative for availability. Drawer I Drawer 4 Drawer 2 Drawer 3 Copyright O Siemens Healthcare GmbH 2020 Drawer 4 Drawer I Unlock Front Cover LWock Frmt Cover Unlock Cover Confirm Cal-QC Storage Content Global Siemens Headquarters Global Siemens HeoLthineers Headquarters Front Cover Henkestr. 127 LWock Frmt Cover Unlock Front Cover Cover Unlock Cover Lock Back Cover 91052 Erlangen, Germany Back Cover Drawer 3 Phone: +49 913184 0 Unlock Back Back Lock Back Cover siemens.com/healthineers Tuesday, February 2B, 2017 101441 AM Tuesday, February 2B, 2017 AM Tuesday, 2B, Tuesday, February 2B, 2017 10 1441 AM Monday, Apri IS, 2019 Monday, April IS, 2019
- Atelica
- Attellica
- Attelica